00
Correct
00
Incorrect
00 : 00 : 00
Session Time
00 : 00
Average Question Time ( Secs)
  • Question 1 - A transport ventilator is powered by an air/oxygen mix using a full oxygen...

    Incorrect

    • A transport ventilator is powered by an air/oxygen mix using a full oxygen cylinder (class CD) with an internal capacity of 2 litres, and pressure of 23,000 kPa, with a gas flow of 4 litres/minute.
      The ventilator also has a control resulting in an additional gas consumption of 1 litre/minute.

      How long will it take for the cylinder to empty?

      Your Answer: 11.5 minutes

      Correct Answer: 92 minutes

      Explanation:

      The Drager Oxylog® 1000 is a pneumatically powered, time-dependent, volume-titrated emergency ventilator with a pressure limit. It is compatible with CD cylinder oxygen. The CD cylinder is a strong and lightweight cylinder usually composed of aluminium or Kevlar. The internal cylinder volume is 2 litres, and the pressure of a full cylinder is 230 bar. The volume of the full cylinder is determined by applying Boyle’s law: P1 × V1 = P2 × V2

      Where:
      P1= pressure of a full cylinder (230 bar)
      V1= volume of oxygen at that pressure (2 litres)
      P2= final pressure (1 bar), and
      V2= volume of oxygen in the full cylinder.

      Substituting values into the equation:

      230 × 2 = 1 x V2
      V2 = 460 litres. The flow of fresh gas is 4 litres/minute + 1 litre/minute required by the control, making a total of 5 litres/minute. The amount of time it takes for the cylinder to empty would be the total volume of oxygen in the full cylinder divided by the amount of oxygen expelled per minute: 460/5 = 92, meaning it would take 92 minutes for the cylinder to empty.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      1030.1
      Seconds
  • Question 2 - Regarding the emergency oxygen flush, which is true? ...

    Incorrect

    • Regarding the emergency oxygen flush, which is true?

      Your Answer: Delivers oxygen that has bypassed the vaporisers but not the flowmeters

      Correct Answer: May lead to awareness if used inappropriately

      Explanation:

      When the emergency oxygen flush is pressed, 100% oxygen is supplied from the common gas outlet. This gas bypasses BOTH flowmeters and vaporisers. The flow of oxygen is usually 45 l/min at a PRESSURE OF 400 kPa.

      There is an increased risk of pulmonary barotrauma when the emergency flush is pressed, especially when anaesthetising paediatric patients.

      The inappropriate use of the flush causes dilution of anaesthetic gases and this increases the possibility of anaesthetic awareness .

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      33
      Seconds
  • Question 3 - From the following electromagnetic waves, which one has the shortest wavelength? ...

    Incorrect

    • From the following electromagnetic waves, which one has the shortest wavelength?

      Your Answer:

      Correct Answer: X rays

      Explanation:

      Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.

      Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.

      Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 4 - For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using...

    Incorrect

    • For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using 100% oxygen and a fresh gas flow equal to the patient's minute ventilation.

      Which would be the most suitable choice of anaesthetic breathing system in this situation?

      Your Answer:

      Correct Answer: Mapleson A system

      Explanation:

      The Mapleson A (Magill) and coaxial version of the Mapleson A system (Lack circuit) are more efficient for spontaneous breathing than any of the other Mapleson circuits. The fresh gas flow (FGF) required to prevent rebreathing is slightly greater than the alveolar minute ventilation (4-5 litres/minute). This is delivered to the patient through the outer coaxial tube and exhaust gases are moved to the scavenging system through the inner tube. In the Lack circuit, the expiratory valve is located close to the common gas outlet away from the patient end. This is the main advantage of the Lack circuit over the Mapleson A circuit.

      The Mapleson E circuit is a modification of the Ayres T piece and the FGF required to prevent rebreathing is 1.5-2 times the patient’s minute volume.

      The Bain circuit is the coaxial version of the Mapleson D circuit.

      The FGF for spontaneous respiration to avoid rebreathing is 160-200 ml/kg/minute.

      The FGF for controlled ventilation to avoid rebreathing is 70-100 ml/kg/min.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 5 - What feature is found in the flowmeters of modern anaesthetic machines ? ...

    Incorrect

    • What feature is found in the flowmeters of modern anaesthetic machines ?

      Your Answer:

      Correct Answer: The use of stannic oxide increases the accuracy of flowmeters

      Explanation:

      Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.

      Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.

      Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.

      Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 6 - Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency.

    For...

    Incorrect

    • Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency.

      For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?

      Your Answer:

      Correct Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts

      Explanation:

      The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.

      These are the following:

      ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
      EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
      EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.

      Prior to display, these small signals will need to be amplified and processed further.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 7 - Which is the most appropriate statement describing the function of flowmeters? ...

    Incorrect

    • Which is the most appropriate statement describing the function of flowmeters?

      Your Answer:

      Correct Answer: Constant pressure, variable orifice - Heidbrink flowmeters

      Explanation:

      There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 8 - Which statement is correct about the Mapleson anaesthetic breathing circuits? ...

    Incorrect

    • Which statement is correct about the Mapleson anaesthetic breathing circuits?

      Your Answer:

      Correct Answer: Mapleson A is most efficient for spontaneous ventilation

      Explanation:

      Mapleson breathing system (or circuit) analysed five different arrangements of components of the breathing system:
      Mapleson A – It is the most efficient for spontaneous respiration. The flow of fresh gas required is 70-85 ml/kg/min, i.e., approximately 5-6 lit./min fresh gas flow for an average adult.
      Mapleson B and C – inefficient for both SV and PPV; requires gas flow of two to three times minute volume (100 ml/kg/min). Not commonly used but category C may be used for emergency resuscitation.
      Mapleson D – efficient for PPV at gas flow equivalent to patient’s minute volume; the Bain’s circuit is a coaxial version of the Mapleson D
      Mapleson E and F – for paediatric use; requires gas flow at two to three times the patient’s minute volume. The Mapleson F consists of an open-ended reservoir bag (Jackson-Rees modification).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 9 - Regarding a paramagnetic oxygen analyser, the following statements are TRUE: ...

    Incorrect

    • Regarding a paramagnetic oxygen analyser, the following statements are TRUE:

      Your Answer:

      Correct Answer: Utilises null deflection

      Explanation:

      The electrons in the outer shell of an oxygen molecule are unpaired, thus it has paramagnetic properties and is attracted into a magnetic field.

      It utilizes null deflection -True
      Null deflection is a crucial principle in paramagnetic analysers (reflected beam of light on two photocells) which gives very accurate results (typically 0.1%).

      It can be used to measure the concentration of diamagnetic gases – False
      Since most other gases are weakly diamagnetic they are repelled by a magnetic field (nitric oxide is also paramagnetic).

      Can measure gases dissolved in the blood – False
      For accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. Therefore, they are unsuitable to measure gases dissolved in blood.

      Does not require calibration – False
      As with most measurement instruments paramagnetic analysers must be calibrated before use.

      E) The readings are unaffected by water vapour – False
      Water vapour affects the readings hence for accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. That is why they are unsuitable to measure dissolved blood gases.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 10 - Dinamap is an automated blood pressure monitoring device. Which of these statements best...

    Incorrect

    • Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?

      Your Answer:

      Correct Answer: The cuff should be positioned at the same level as the heart

      Explanation:

      Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.

      The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.

      The manual BP device is still the gold standard for BP measurement and monitoring.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 11 - Bacteria and viruses that are smaller than 0.1 ?m in diameter can be...

    Incorrect

    • Bacteria and viruses that are smaller than 0.1 ?m in diameter can be filtered out using heat and moisture exchanger (HME) with a typical pore size 0.2 ?m.

      Choose the most appropriate mechanisms of particle capture for most bacteria and viruses.

      Your Answer:

      Correct Answer: Diffusion

      Explanation:

      Warming, humidifying, and filtering inspired anaesthetic gases is done by heat and moisture exchangers (HME) and breathing system filters. They are made of glass fibres materials and are supported by a sturdy frame. Pleating increases the surface area to reduce resistance to air flow and boost efficiency.

      Filters’ effectiveness is determined by the amount and size of particles they keep out of the patient’s airway. The efficiency of filters might be classified as 95, 99.95, or 99.97 percent. Pores with a diameter of 0.2 µm are common. The following are examples of typical particle sizes:
      Red blood cell – 5 µm
      Lymphocyte – 5-8 µm
      Viruses – 0.02-0.3 µm
      Bacteria – 0.5-1 µm
      Depending on particle size, gas flow speed, and charge, particles are collected via a number of processes. Mechanical sieve, interception, diffusion, electrostatic filtration, and inertial impaction are some of the options:

      Sieve:
      The diameter of the particle the filter is supposed to collect is smaller than the apertures of the filter’s fibres.

      Interception:
      When a particle following a gas streamline approaches a fibre within one radius of itself, it becomes attached and captured.
      Diffusion:

      A particle’s random (Brownian) zig-zag path or motion causes it to collide with a fibre.
      By attracting and capturing a particle from within the gas flow, it generates a lower-concentration patch within the gas flow into which another particle diffuses, only to be captured. At low gas velocities and with smaller particles (0.1µm diameter), this is more common.

      Electrostatic:

      These filters use large diameter fibre media and rely on electrostatic charges to improve fine particle removal effectiveness.

      Impaction due to inertia:

      When a particle is too large to respond fast to abrupt changes in streamline direction near a filter fibre, this happens. Because of its inertia, the particle will continue on its original course and collide with the filter fibre. When high gas velocities and dense fibre packing of the filter media are present, this sort of filtration mechanism is most prevalent.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 12 - A strain gauge operates on which of the following principles? ...

    Incorrect

    • A strain gauge operates on which of the following principles?

      Your Answer:

      Correct Answer: Varying resistance of a wire when stretched

      Explanation:

      The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.

      A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.

      Changes in current flow through the resistor are amplified and displayed as a pressure change measure.

      A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 13 - Which of the following is true about the pulse oximeter? ...

    Incorrect

    • Which of the following is true about the pulse oximeter?

      Your Answer:

      Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation

      Explanation:

      The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.

      The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.

      An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.

      The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 14 - A 240 volt alternating current (AC) socket from a wall is used to...

    Incorrect

    • A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator.

      Name the electrical component that converts AC to DC.

      Your Answer:

      Correct Answer: Rectifier

      Explanation:

      There are two types of defibrillators
      AC defibrillator
      DC defibrillator

      AC defibrillator,
      consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.

      DC defibrillator,
      consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.

      Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.

      Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.

      Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.

      In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 15 - A measuring system's response to change is complex, yet it can be mathematically modelled.

    Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?

    ...

    Incorrect

    • A measuring system's response to change is complex, yet it can be mathematically modelled.

      Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?

      Your Answer:

      Correct Answer: Dynamic second-order response

      Explanation:

      The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
      The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:

      Zero-order:
      Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.

      First-order:
      Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.

      Second-order
      Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.

      Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 16 - With respect to the peripheral nerve stimulators, which one is used to perform...

    Incorrect

    • With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?

      Your Answer:

      Correct Answer:

      Explanation:

      The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.

      There are just 2 leads (not 3); one for the skin and other for the needle.

      Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.

      If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.

      Insulated needles have improved the block success rate, as the current is only conducting through needle tip.

      Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 17 - A sevoflurane vaporiser with a 2 percent setting and a 200 kPa ambient...

    Incorrect

    • A sevoflurane vaporiser with a 2 percent setting and a 200 kPa ambient pressure is used.

      At this pressure, which of the following options best represents vaporiser output?

      Your Answer:

      Correct Answer: The output is 1% because the saturated pressure of sevoflurane is unaffected by ambient pressure

      Explanation:

      Ambient pressure has no effect on a volatile agent’s saturated vapour pressure (SVP). At a temperature of 20°C, the SVP of sevoflurane is approximately 21 kPa, or 21% of atmospheric pressure (100 kPa).

      The SVP of sevoflurane remains the same when the ambient pressure is doubled to 200 kPa, but the output of the vaporiser is halved, now 21 percent of 200 kPa, equalling 10.5 percent. The vaporiser’s output has increased to 1%, but the partial pressure output has remained unchanged. The splitting ratio will not change because it is determined by temperature changes.

      Calculations can be made as follows:

      Vaporizer output % (ambient pressure) = % volatile (calibrated) x 100 kPa calibrated pressure/ambient pressure
      2% = 2% (dialled) × 100/100
      2% of 100 = 2 kPa

      Altitude, pressure 50 kPa
      4% = 2% (dialled) × 100/50
      4% of 50 = 2 kPa

      High pressure at 200 kPa
      1% = 2% (dialled) × 100/200
      1% of 200 = 2 kPa

      Sevoflurane has a boiling point of 58°C and, unlike desflurane (which has a boiling point of 22.8°C), does not need to be heated and pressurised with a Tec 6 vaporiser.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 18 - A laser is a device that stimulates atoms or molecules to emit light...

    Incorrect

    • A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment.

      Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?

      Your Answer:

      Correct Answer: Wearing laser protective goggles

      Explanation:

      Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.

      Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.

      Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.

      Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage. Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams. The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury. Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions. Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
      Using short bursts to reduce energy is also not correct as it would still be harmful to eye.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 19 - When compared to unipolar diathermy, which of the following is more specific to...

    Incorrect

    • When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?

      Your Answer:

      Correct Answer: Has a power output of up to 140 joules per second

      Explanation:

      Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).

      The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.

      A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.

      High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).

      The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 20 - A 74-year-old with a VVI pacemaker is undergoing a hip replacement.

    Which of the...

    Incorrect

    • A 74-year-old with a VVI pacemaker is undergoing a hip replacement.

      Which of the following is most likely to predispose him to an electrical hazard?

      Your Answer:

      Correct Answer: Use of cutting unipolar diathermy

      Explanation:

      A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).

      EMI has the potential to cause the following: Inhibition of pacing
      Asynchronous pacing
      Reset to backup mode
      Myocardial burns, and
      Trigger VF.

      Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.

      Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.

      Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.

      Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).

      A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 21 - Regarding the Manley MP3 ventilator, which statement is true? ...

    Incorrect

    • Regarding the Manley MP3 ventilator, which statement is true?

      Your Answer:

      Correct Answer: Is a minute volume divider

      Explanation:

      It’s a minute volume divider – True
      The Manley MP3 ventilator is classed as a minute volume divider. The entire fresh gas flow or minute volume is delivered to the patient, having been divided into readily set tidal volumes.

      Can efficiently ventilate patients with poor pulmonary compliance – False
      Ventilating patients with poor pulmonary compliance is not easily achieved, which makes it an unsuitable ventilator for a modern ICU.

      Can generate tidal volume up to 1500ml – False
      It can generate tidal volumes up to 1000 ml (not 1500 ml), and the inflation pressure can be adjusted by sliding a weight along a rail.

      Functions like a Mapleson A system during spontaneous ventilation – False
      The ventilator functions like a Mapleson D breathing system (not Mapleson A) during spontaneous ventilation.

      Has three sets of bellows – False
      The fresh gas flow drives the ventilator which allows rapid detection of gas supply failure. It has two sets of bellows (not three) and three unidirectional valves.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 22 - International colour coding is used on medical gas cylinders. Other characteristics also play...

    Incorrect

    • International colour coding is used on medical gas cylinders. Other characteristics also play a role in determining the gas's identity within a cylinder.

      Which of the following options best describes a cylinder containing analgesics for obstetrics?

      Your Answer:

      Correct Answer: Blue body, blue/white shoulder, full cylinder; 13700 KPa, gas mixture, requires a dual stage pressure regulator

      Explanation:

      The body of the Entonox cylinder is usually blue (occasionally white), with blue and white shoulders. Entonox contains a 50:50 mixture of oxygen and nitrous oxide, with a full cylinder pressure of 13700 KPa (137 bar). The cylinder is equipped with a two-stage pressure regulator for safe operation.

      The cylinder body and shoulder of nitrous oxide are (French) blue.

      In today’s anaesthetic workstations, carbon dioxide cylinders are no longer used.

      The body of an oxygen cylinder is black, with a white shoulder.

      The white Heliox (21 percent oxygen and 79 percent helium) cylinder has a brown and white shoulder. The administration of this gas mixture, which is less dense than air, is used to reduce turbulence (stridor) of inspiratory flow in patients with upper airway obstruction.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 23 - Which of the following is true regarding a laryngoscope? ...

    Incorrect

    • Which of the following is true regarding a laryngoscope?

      Your Answer:

      Correct Answer: The Wisconsin and Seward are examples of straight blade laryngoscopes

      Explanation:

      Direct laryngoscopy are performed using laryngoscopes and they can be classed according to the shape of the blade as curved or straight.

      Miller, Soper, Wisconsin and Seward are examples of straight blade laryngoscopes. Straight blades are commonly used for intubating neonates and infants but can be used in adults too.

      The tip of the miller blade is advanced over the epiglottis to the tracheal entrance then lifted in order to view the vocal cords.

      The RIGHT-SIDED Macintosh blade is used in adults while the left-sided blade may be used in conditions that make intubation with standard blade difficult e.g. facial deformities.

      The McCoy laryngoscope is based on the STANDARD MACINTOSH blade not Robertshaw’s. It has a lever operated hinged tip, which improves the view during laryngoscopy.

      Polio blade is mounted at an angle of 120-135 degrees to the handle. Originally designed for use during the polio epidemic ​in intubation patients within iron lung ventilators, it is now useful in patients with conditions like breast hypertrophy, barrel chest, and restricted neck mobility.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 24 - All of the following statements about pH electrode are incorrect except: ...

    Incorrect

    • All of the following statements about pH electrode are incorrect except:

      Your Answer:

      Correct Answer: A semi-permeable membrane reduces protein contamination

      Explanation:

      Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.

      The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.

      Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.

      Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 25 - Which statement most accurately describes the action of increasing the gain in ultrasound...

    Incorrect

    • Which statement most accurately describes the action of increasing the gain in ultrasound imaging?

      Your Answer:

      Correct Answer: Amplifies the returning signal

      Explanation:

      A higher frequency ultrasound comes with a better resolution of the digital image. Ultrasound with a frequency of 15 MHz is best used in imaging of superficial organs such as the thyroid gland, muscles, tendons and breasts whereas deep organs are better imaged using a lower frequency of 2-7MHz because of its ability for deeper penetration but lower resolution. These low frequency probes are also used to diagnose ascites, pleural effusions or can be used in echocardiography.

      The US probe emits and then absorbs a reflected wave. Similar to brightness control, increasing the gain will amplify the return signal which is then attenuated by the tissue. This increases the signal to noise ratio.
      A high frame rate, which basically means the number of times an image is updated onto the screen per second, improves the resolution of a moving 3D image which has become more accurate as the computing power has increased.

      Widening of the image field can be obtained by altering the penetration depth which is obtained by changing the frequency of the US beam

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 26 - Regarding laminar gas flow, which of the following options has the most influence...

    Incorrect

    • Regarding laminar gas flow, which of the following options has the most influence on laminar flow?

      Your Answer:

      Correct Answer: Diameter of tube

      Explanation:

      Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:

      1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
      2. Few directional changes
      3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
      4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
      5. Double flow by doubling pressure as long as the flow pattern remains laminar

      Poiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 27 - Regarding tracheal tubes, which of the following statements are true? ...

    Incorrect

    • Regarding tracheal tubes, which of the following statements are true?

      Your Answer:

      Correct Answer: Uncuffed RAE tubes have two Murphy eyes

      Explanation:

      Tracheal tubes are made of either disposable plastic or reusable red rubber.

      The tube size refers to the internal diameter (ID) in mm which is marked on the outside of the tube (some manufacturers mark the external diameter on the outside).

      Plastic tubes have a radiopaque line spanning the entire length of the tube, which allows their position to be identified on x-rays. The bevel located at the end of the tube is left-facing and oval in shape, which improves the view of the vocal cords during intubation.

      Oxford tubes are L-shaped and have a bevel that faces posteriorly. They have thick walls that increase the external diameter, making for a wider internal diameter.

      RAE (Ring, Adair, and Elwyn) tubes are preformed and can either be north or south facing and cuffed or uncuffed. The cuffed RAE tubes have one Murphy eye, whereas the uncuffed has two Murphy eyes. Uncuffed tubes are primarily used in paediatric anaesthesia and the two Murphy eyes ensure adequate ventilation- should the tube be too long.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 28 - A current flows through a simple electric circuit.

    Which of the following electrical component...

    Incorrect

    • A current flows through a simple electric circuit.

      Which of the following electrical component configurations has the greatest potential difference?

      Your Answer:

      Correct Answer: Two 5 ohm resistors in series with a passing current of 10 ampere

      Explanation:

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 29 - When the volume of gas is measured, its value should be supported by...

    Incorrect

    • When the volume of gas is measured, its value should be supported by a suitable abbreviation. From the following options, what does the STPD stand for?

      Your Answer:

      Correct Answer: Standard Temperature and Pressure, Dry

      Explanation:

      Gas is composed of large numbers of molecules moving in random directions, separated by distances. They undergo perfectly elastic collisions with each other and the walls of a container and transfer kinetic energy in form of heat. These assumptions bring the characteristics of gases within the range and reasonable approximation to a real gas, particularly how any change in temperature and pressure affect the behaviour of gas. According to different theories and laws proposed, mathematical equations are derived to calculate the volume of gas, also different abbreviations are being used according to given conditions. The abbreviations used are ATP, BTPS, and STPD.
      ATP stands for ambient temperature and barometric pressure, it is used to describe the conditions under which volume of gas is measured.
      BTPS stands for body temperature and pressure saturated with water vapor. These are conditions under which volume of gas exist and all results of lung volume determination should be quoted at BTPS.
      STPD stands for standard temperature and pressure, dry (0C and 760 mm Hg). These are the conditions that are used to describe quantities of individual gases exchanged in the lungs.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 30 - A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted...

    Incorrect

    • A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted in the surgery department for an elective total thyroidectomy with radical neck dissection. The operation is expected to last for 10 hours.

      Which of the following is the most suitable humidifier to use in an anaesthetic circuit for this case?

      Your Answer:

      Correct Answer: Heat and moisture exchanger (HME)

      Explanation:

      Adequate humidification is vital to maintain homeostasis of the airway. Heat and moisture exchangers conserve some of the exhaled water, heat and return them to inspired gases. Many heat and moisture exchangers also perform bacterial/viral filtration and prevent inhalation of small particles. Heat and moisture exchangers are also called condenser humidifier, artificial nose, etc. Most of them are disposable devices with exchanging medium enclosed in a plastic housing. For adult and paediatric age group different dead space types are available. Heat and moisture exchangers are helpful during anaesthesia and ventilatory breathing system. To reduce the damage of the upper respiratory tract through cooling and dehydration inspiratory air can be heated and humidified, thus preventing the serious complications. Moreover, they are the most appropriate humidification devices used for routine anaesthesia.

      Gases can be bubbled through water to increase humidity. Passing gas through water at room temperature causes the gas to cool due to latent heat of vaporisation. The water bath can be heated. This improves the efficiency of the device and also reduces the incidence of bacterial colonisation.

      Nebulisers use a venturi system which employs the Bernoulli effect. A gas at high flow passes through a constriction causing the gas to accelerate, reducing its potential energy allowing other gases or liquids to be entrained. This can include medications or in the case of humidification, water vapour. The size of the water droplet produced by nebulisation determines where in the airway it is deposited. Standard nebulisers produced droplets of 4 microns in diameter and these are deposited in the upper airway and trachea. Efficacy can be improved by passing the droplets over an anvil which further reduces particle size. The most efficient form of nebuliser is the ultrasonic nebuliser. Here a transducer immersed in water and vibrated at a frequency of 3MHz produces1-2micron droplets. These particles easily reach the bronchioles and provide excellent humidification.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds

SESSION STATS - PERFORMANCE PER SPECIALTY

Anaesthesia Related Apparatus (1/2) 50%
Passmed