  • Question 1 - Which of the following is correct for gas pipeline pressure? ...

    Incorrect

    • Which of the following is correct for gas pipeline pressure?

      Your Answer: 4 bar for oxygen

      Correct Answer:

      Explanation:

      Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.

      Carbon dioxide and nitric oxide are usually only supplied in cylinders.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      13.9
      Seconds
  • Question 2 - A 74-year-old with a VVI pacemaker is undergoing a hip replacement.

    Which of the...

    Correct

    • A 74-year-old with a VVI pacemaker is undergoing a hip replacement.

      Which of the following is most likely to predispose him to an electrical hazard?

      Your Answer: Use of cutting unipolar diathermy

      Explanation:

      A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).

      EMI has the potential to cause the following: Inhibition of pacing
      Asynchronous pacing
      Reset to backup mode
      Myocardial burns, and
      Trigger VF.

      Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.

      Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.

      Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.

      Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).

      A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      24.3
      Seconds
  • Question 3 - With respect to the peripheral nerve stimulators, which one is used to perform...

    Incorrect

    • With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?

      Your Answer: Prior to injecting the local anaesthetic the ideal current is 1-2 mili amperes at a frequency of 1-2 Hz

      Correct Answer:

      Explanation:

      The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.

      There are just 2 leads (not 3); one for the skin and other for the needle.

      Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.

      If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.

      Insulated needles have improved the block success rate, as the current is only conducting through needle tip.

      Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      27.3
      Seconds
  • Question 4 - Which is the most appropriate statement describing the function of flowmeters? ...

    Incorrect

    • Which is the most appropriate statement describing the function of flowmeters?

      Your Answer: Variable pressure, variable orifice - water depression flowmeter

      Correct Answer: Constant pressure, variable orifice - Heidbrink flowmeters

      Explanation:

      There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      46.4
      Seconds
  • Question 5 - Among the following, which statement is true regarding electrical safety in an operation...

    Incorrect

    • Among the following, which statement is true regarding electrical safety in an operation theatre?

      Your Answer: Class III equipment are double-insulated

      Correct Answer: The higher the frequency of the current the less risk to the patient

      Explanation:

      The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.

      Electricity can cause morbidity or mortality by one of the following ways:
      (i) electrocution
      (ii) burns
      (iii) ignition of a flammable material, causing a fire or explosion.

      Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.

      Option A: The higher the frequency, the less effects of electrocution on the body.

      Option B & D: Equipment can be classified in classes and types.
      The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
      The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
      Type B:
      can be class I, II or III but the maximum leakage current must not exceed 100 µA. It is therefore not suitable for direct connection to the heart.
      Type BF
      Similar to type B, but uses an isolated (or floating) circuit.
      Type CF
      Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 ?A is sufficient to cause VF.

      Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
      1 mA – tingling pain
      5 mA – pain
      15 mA – tonic muscular contraction
      50 mA – respiratory muscle paralysis
      75 mA – ventricular fibrillation.

      Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      11
      Seconds
  • Question 6 - Which medical gas cylinders have the correct colour codes? ...

    Incorrect

    • Which medical gas cylinders have the correct colour codes?

      Your Answer:

      Correct Answer: Oxygen cylinders have a black body with white shoulders

      Explanation:

      The following are the colour codes for medical gas cylinders:

      Oxygen cylinder has a dark body with white shoulders.

      Nitrous oxide is French blue. Air encompasses a grey body with dark and white quarters on the shoulders.

      Entonox contains a French blue body with white and blue quarters on the shoulders.

      Carbon dioxide barrels are grey in colour.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 7 - Regarding tracheal tubes, which of the following statements are true? ...

    Incorrect

    • Regarding tracheal tubes, which of the following statements are true?

      Your Answer:

      Correct Answer: Uncuffed RAE tubes have two Murphy eyes

      Explanation:

      Tracheal tubes are made of either disposable plastic or reusable red rubber.

      The tube size refers to the internal diameter (ID) in mm which is marked on the outside of the tube (some manufacturers mark the external diameter on the outside).

      Plastic tubes have a radiopaque line spanning the entire length of the tube, which allows their position to be identified on x-rays. The bevel located at the end of the tube is left-facing and oval in shape, which improves the view of the vocal cords during intubation.

      Oxford tubes are L-shaped and have a bevel that faces posteriorly. They have thick walls that increase the external diameter, making for a wider internal diameter.

      RAE (Ring, Adair, and Elwyn) tubes are preformed and can either be north or south facing and cuffed or uncuffed. The cuffed RAE tubes have one Murphy eye, whereas the uncuffed has two Murphy eyes. Uncuffed tubes are primarily used in paediatric anaesthesia and the two Murphy eyes ensure adequate ventilation- should the tube be too long.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 8 - When the volume of gas is measured, its value should be supported by...

    Incorrect

    • When the volume of gas is measured, its value should be supported by a suitable abbreviation. From the following options, what does the STPD stand for?

      Your Answer:

      Correct Answer: Standard Temperature and Pressure, Dry

      Explanation:

      Gas is composed of large numbers of molecules moving in random directions, separated by distances. They undergo perfectly elastic collisions with each other and the walls of a container and transfer kinetic energy in form of heat. These assumptions bring the characteristics of gases within the range and reasonable approximation to a real gas, particularly how any change in temperature and pressure affect the behaviour of gas. According to different theories and laws proposed, mathematical equations are derived to calculate the volume of gas, also different abbreviations are being used according to given conditions. The abbreviations used are ATP, BTPS, and STPD.
      ATP stands for ambient temperature and barometric pressure, it is used to describe the conditions under which volume of gas is measured.
      BTPS stands for body temperature and pressure saturated with water vapor. These are conditions under which volume of gas exist and all results of lung volume determination should be quoted at BTPS.
      STPD stands for standard temperature and pressure, dry (0C and 760 mm Hg). These are the conditions that are used to describe quantities of individual gases exchanged in the lungs.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 9 - Which of the following is true about the pulse oximeter? ...

    Incorrect

    • Which of the following is true about the pulse oximeter?

      Your Answer:

      Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation

      Explanation:

      The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.

      The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.

      An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.

      The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 10 - A strain gauge operates on which of the following principles? ...

    Incorrect

    • A strain gauge operates on which of the following principles?

      Your Answer:

      Correct Answer: Varying resistance of a wire when stretched

      Explanation:

      The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.

      A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.

      Changes in current flow through the resistor are amplified and displayed as a pressure change measure.

      A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 11 - What feature is found in the flowmeters of modern anaesthetic machines ? ...

    Incorrect

    • What feature is found in the flowmeters of modern anaesthetic machines ?

      Your Answer:

      Correct Answer: The use of stannic oxide increases the accuracy of flowmeters

      Explanation:

      Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.

      Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.

      Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.

      Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 12 - Which statement most accurately describes the action of increasing the gain in ultrasound...

    Incorrect

    • Which statement most accurately describes the action of increasing the gain in ultrasound imaging?

      Your Answer:

      Correct Answer: Amplifies the returning signal

      Explanation:

      A higher frequency ultrasound comes with a better resolution of the digital image. Ultrasound with a frequency of 15 MHz is best used in imaging of superficial organs such as the thyroid gland, muscles, tendons and breasts whereas deep organs are better imaged using a lower frequency of 2-7MHz because of its ability for deeper penetration but lower resolution. These low frequency probes are also used to diagnose ascites, pleural effusions or can be used in echocardiography.

      The US probe emits and then absorbs a reflected wave. Similar to brightness control, increasing the gain will amplify the return signal which is then attenuated by the tissue. This increases the signal to noise ratio.
      A high frame rate, which basically means the number of times an image is updated onto the screen per second, improves the resolution of a moving 3D image which has become more accurate as the computing power has increased.

      Widening of the image field can be obtained by altering the penetration depth which is obtained by changing the frequency of the US beam

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 13 - A current flows through a simple electric circuit.

    Which of the following electrical component...

    Incorrect

    • A current flows through a simple electric circuit.

      Which of the following electrical component configurations has the greatest potential difference?

      Your Answer:

      Correct Answer: Two 5 ohm resistors in series with a passing current of 10 ampere

      Explanation:

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 14 - The equipment used for patient monitoring in theatre and intensive care settings have...

    Incorrect

    • The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients.
      Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?

      Your Answer:

      Correct Answer: II (CF)

      Explanation:

      Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
      The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
      II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
      III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
      Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
      B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
      BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
      CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 15 - Which statement is correct about the Mapleson anaesthetic breathing circuits? ...

    Incorrect

    • Which statement is correct about the Mapleson anaesthetic breathing circuits?

      Your Answer:

      Correct Answer: Mapleson A is most efficient for spontaneous ventilation

      Explanation:

      Mapleson breathing system (or circuit) analysed five different arrangements of components of the breathing system:
      Mapleson A – It is the most efficient for spontaneous respiration. The flow of fresh gas required is 70-85 ml/kg/min, i.e., approximately 5-6 lit./min fresh gas flow for an average adult.
      Mapleson B and C – inefficient for both SV and PPV; requires gas flow of two to three times minute volume (100 ml/kg/min). Not commonly used but category C may be used for emergency resuscitation.
      Mapleson D – efficient for PPV at gas flow equivalent to patient’s minute volume; the Bain’s circuit is a coaxial version of the Mapleson D
      Mapleson E and F – for paediatric use; requires gas flow at two to three times the patient’s minute volume. The Mapleson F consists of an open-ended reservoir bag (Jackson-Rees modification).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 16 - Regarding the Manley MP3 ventilator, which statement is true? ...

    Incorrect

    • Regarding the Manley MP3 ventilator, which statement is true?

      Your Answer:

      Correct Answer: Is a minute volume divider

      Explanation:

      It’s a minute volume divider – True
      The Manley MP3 ventilator is classed as a minute volume divider. The entire fresh gas flow or minute volume is delivered to the patient, having been divided into readily set tidal volumes.

      Can efficiently ventilate patients with poor pulmonary compliance – False
      Ventilating patients with poor pulmonary compliance is not easily achieved, which makes it an unsuitable ventilator for a modern ICU.

      Can generate tidal volume up to 1500ml – False
      It can generate tidal volumes up to 1000 ml (not 1500 ml), and the inflation pressure can be adjusted by sliding a weight along a rail.

      Functions like a Mapleson A system during spontaneous ventilation – False
      The ventilator functions like a Mapleson D breathing system (not Mapleson A) during spontaneous ventilation.

      Has three sets of bellows – False
      The fresh gas flow drives the ventilator which allows rapid detection of gas supply failure. It has two sets of bellows (not three) and three unidirectional valves.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 17 - A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted...

    Incorrect

    • A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted in the surgery department for an elective total thyroidectomy with radical neck dissection. The operation is expected to last for 10 hours.

      Which of the following is the most suitable humidifier to use in an anaesthetic circuit for this case?

      Your Answer:

      Correct Answer: Heat and moisture exchanger (HME)

      Explanation:

      Adequate humidification is vital to maintain homeostasis of the airway. Heat and moisture exchangers conserve some of the exhaled water, heat and return them to inspired gases. Many heat and moisture exchangers also perform bacterial/viral filtration and prevent inhalation of small particles. Heat and moisture exchangers are also called condenser humidifier, artificial nose, etc. Most of them are disposable devices with exchanging medium enclosed in a plastic housing. For adult and paediatric age group different dead space types are available. Heat and moisture exchangers are helpful during anaesthesia and ventilatory breathing system. To reduce the damage of the upper respiratory tract through cooling and dehydration inspiratory air can be heated and humidified, thus preventing the serious complications. Moreover, they are the most appropriate humidification devices used for routine anaesthesia.

      Gases can be bubbled through water to increase humidity. Passing gas through water at room temperature causes the gas to cool due to latent heat of vaporisation. The water bath can be heated. This improves the efficiency of the device and also reduces the incidence of bacterial colonisation.

      Nebulisers use a venturi system which employs the Bernoulli effect. A gas at high flow passes through a constriction causing the gas to accelerate, reducing its potential energy allowing other gases or liquids to be entrained. This can include medications or in the case of humidification, water vapour. The size of the water droplet produced by nebulisation determines where in the airway it is deposited. Standard nebulisers produced droplets of 4 microns in diameter and these are deposited in the upper airway and trachea. Efficacy can be improved by passing the droplets over an anvil which further reduces particle size. The most efficient form of nebuliser is the ultrasonic nebuliser. Here a transducer immersed in water and vibrated at a frequency of 3MHz produces1-2micron droplets. These particles easily reach the bronchioles and provide excellent humidification.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 18 - Which statement is correct concerning breathing systems? ...

    Incorrect

    • Which statement is correct concerning breathing systems?

      Your Answer:

      Correct Answer: The reservoir bag can limit the pressure in the breathing system to about 40 cm of water

      Explanation:

      Mapleson classified breathing systems into A, B, C, D and E. Jackson-Rees subsequently modified the Mapleson E by adding a double-ended bag to the end of the reservoir tubing, creating the Mapleson F. A Mapleson E or T-piece does not have a reservoir bag.

      A Mapleson A system is a very efficient system for use during spontaneous ventilation. However, it is not suitable for use with patients less than 25 kg, due to the increased dead space at the distal / patient end. This system can be modified into a Lack system or coaxial Mapleson A, where the fresh gas flows through an outer tube (30 mm) and exhaled gases flow through the inner tube (14 mm).

      The adjustable pressure limiting valve (APL) or expiratory valve allows exhaled gas and excess fresh gas to leave the breathing system. It is a one-way, adjustable spring-loaded valve, and gases escape when the pressure in the system exceeds the valve opening pressure. During spontaneous ventilation a pressure of less than 1 cm of water (0.1 kPa) is needed when the valve is in the open position (not 2 cm of H2O).

      The reservoir bag is highly compliant and when over inflated, the rubber bag can limit the pressure in the system to about 40 cm of H2O.

      This is due to the law of Laplace, which states that the pressure will fall as the radius of the bag increases:

      Pressure = 2 x tension/radius.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 19 - All of the following statements about pH electrode are incorrect except: ...

    Incorrect

    • All of the following statements about pH electrode are incorrect except:

      Your Answer:

      Correct Answer: A semi-permeable membrane reduces protein contamination

      Explanation:

      Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.

      The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.

      Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.

      Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 20 - You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy...

    Incorrect

    • You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy to rule out appendicitis.

      She has no medical history and does not take any medications on a regular basis. You're going to do a quick sequence induction.

      Which method of preoxygenation is the most effective and efficient?

      Your Answer:

      Correct Answer: Mapleson A circuit with a fresh gas flow of 100 ml/kg

      Explanation:

      Professor Mapleson classified non-rebreathing circuits based on the position of the APL valve, which controls fresh gas flow.

      The Mapleson A (Magill) circuit is most effective in spontaneous breathing, requiring only 70-100 ml/kg (the patient’s minute volume) of fresh gas flow. The patient inhales fresh gas from the reservoir bag and tubing during inspiration. During expiration, the patient adds dead space gas (gas that hasn’t been exchanged) to the tubing and reservoir bag in addition to the fresh gas flow. At the patient’s end, alveolar gas is vented through the APL valve. During the expiratory pause, the fresh gas flow causes more gas to be released.

      The Mapleson A is inefficient during controlled ventilation. Venting occurs during inspiration rather than during the expiratory phase, as it does during spontaneous ventilation. As a result, unless a high fresh gas flow of >20 L/minute is used, alveolar gas is rebreathed.

      During spontaneous ventilation, the Mapleson D circuit is inefficient.

      The oxygen concentration in a Hudson mask is insufficient to allow for adequate pre-oxygenation.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 21 - Many of the processes we deal with in anaesthesia entail a relationship between...

    Incorrect

    • Many of the processes we deal with in anaesthesia entail a relationship between two or more variables.

      Which of the following relationships is a linear relationship?

      Your Answer:

      Correct Answer: The relationship between the junction potential and temperature in a thermocouple

      Explanation:

      Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).

      All of the other connections are non-linear.

      For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.

      A sine wave is the relationship between current and degrees or time from a mains power source.

      A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.

      The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 22 - From the following electromagnetic waves, which one has the shortest wavelength? ...

    Incorrect

    • From the following electromagnetic waves, which one has the shortest wavelength?

      Your Answer:

      Correct Answer: X rays

      Explanation:

      Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.

      Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.

      Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 23 - For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using...

    Incorrect

    • For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using 100% oxygen and a fresh gas flow equal to the patient's minute ventilation.

      Which would be the most suitable choice of anaesthetic breathing system in this situation?

      Your Answer:

      Correct Answer: Mapleson A system

      Explanation:

      The Mapleson A (Magill) and coaxial version of the Mapleson A system (Lack circuit) are more efficient for spontaneous breathing than any of the other Mapleson circuits. The fresh gas flow (FGF) required to prevent rebreathing is slightly greater than the alveolar minute ventilation (4-5 litres/minute). This is delivered to the patient through the outer coaxial tube and exhaust gases are moved to the scavenging system through the inner tube. In the Lack circuit, the expiratory valve is located close to the common gas outlet away from the patient end. This is the main advantage of the Lack circuit over the Mapleson A circuit.

      The Mapleson E circuit is a modification of the Ayres T piece and the FGF required to prevent rebreathing is 1.5-2 times the patient’s minute volume.

      The Bain circuit is the coaxial version of the Mapleson D circuit.

      The FGF for spontaneous respiration to avoid rebreathing is 160-200 ml/kg/minute.

      The FGF for controlled ventilation to avoid rebreathing is 70-100 ml/kg/min.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 24 - Prior to rapid sequence induction of anaesthesia, a man with a BMI of...

    Incorrect

    • Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated.

      Which method of pre-oxygenation with a tight-fitting face mask is the most effective?

      Your Answer:

      Correct Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths

      Explanation:

      This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:

      Functional residual capacity has been reduced (FRC)
      Increased closing capacity (CC)
      Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
      Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.

      Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 25 - Dinamap is an automated blood pressure monitoring device. Which of these statements best...

    Incorrect

    • Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?

      Your Answer:

      Correct Answer: The cuff should be positioned at the same level as the heart

      Explanation:

      Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.

      The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.

      The manual BP device is still the gold standard for BP measurement and monitoring.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 26 - Which of the following statements is TRUE regarding an epidural set? ...

    Incorrect

    • Which of the following statements is TRUE regarding an epidural set?

      Your Answer:

      Correct Answer: 19G Tuohy needles have 0.5 cm markings

      Explanation:

      A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings

      18G Tuohy catheters are generally 9 to 10cm to hub

      Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura

      Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure

      Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings)

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 27 - Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?...

    Incorrect

    • Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?

      Your Answer:

      Correct Answer: Lack circuit

      Explanation:

      Among the breathing circuits, the Lack circuit is the most efficient for spontaneous breathing.

      An outer coaxial tube is present to deliver fresh air; exhaust air is routed to an inner tube, which is then delivered to a scavenging system. An expiratory valve is seen at the patient end, which is an advantage over other circuits. Moreover, the Lack circuit prevents rebreathing slightly greater than the alveolar minute ventilation at 4-5 litres per minute.

      The Bain circuit prevents rebreathing at 160-200ml/kg per minute, and is a co-axial version of the Mapleson D circuit.

      The Mapleson E circuit prevent rebreathing at a fresh gas flow (FGF) of approximately twice the patient’s normal minute volume. A modification of this, the Mapleson F, has a reservoir bag at the opposite end for the FGF. This circuit is appropriate for paediatric patients with a body weight less than 20 kg.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 28 - A transport ventilator is powered by an air/oxygen mix using a full oxygen...

    Incorrect

    • A transport ventilator is powered by an air/oxygen mix using a full oxygen cylinder (class CD) with an internal capacity of 2 litres, and pressure of 23,000 kPa, with a gas flow of 4 litres/minute.
      The ventilator also has a control resulting in an additional gas consumption of 1 litre/minute.

      How long will it take for the cylinder to empty?

      Your Answer:

      Correct Answer: 92 minutes

      Explanation:

      The Drager Oxylog® 1000 is a pneumatically powered, time-dependent, volume-titrated emergency ventilator with a pressure limit. It is compatible with CD cylinder oxygen. The CD cylinder is a strong and lightweight cylinder usually composed of aluminium or Kevlar. The internal cylinder volume is 2 litres, and the pressure of a full cylinder is 230 bar. The volume of the full cylinder is determined by applying Boyle’s law: P1 × V1 = P2 × V2

      Where:
      P1= pressure of a full cylinder (230 bar)
      V1= volume of oxygen at that pressure (2 litres)
      P2= final pressure (1 bar), and
      V2= volume of oxygen in the full cylinder.

      Substituting values into the equation:

      230 × 2 = 1 x V2
      V2 = 460 litres. The flow of fresh gas is 4 litres/minute + 1 litre/minute required by the control, making a total of 5 litres/minute. The amount of time it takes for the cylinder to empty would be the total volume of oxygen in the full cylinder divided by the amount of oxygen expelled per minute: 460/5 = 92, meaning it would take 92 minutes for the cylinder to empty.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 29 - A 240 volt alternating current (AC) socket from a wall is used to...

    Incorrect

    • A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator.

      Name the electrical component that converts AC to DC.

      Your Answer:

      Correct Answer: Rectifier

      Explanation:

      There are two types of defibrillators
      AC defibrillator
      DC defibrillator

      AC defibrillator,
      consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.

      DC defibrillator,
      consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.

      Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.

      Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.

      Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.

      In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 30 - Which of the following is true regarding a laryngoscope? ...

    Incorrect

    • Which of the following is true regarding a laryngoscope?

      Your Answer:

      Correct Answer: The Wisconsin and Seward are examples of straight blade laryngoscopes

      Explanation:

      Direct laryngoscopy are performed using laryngoscopes and they can be classed according to the shape of the blade as curved or straight.

      Miller, Soper, Wisconsin and Seward are examples of straight blade laryngoscopes. Straight blades are commonly used for intubating neonates and infants but can be used in adults too.

      The tip of the miller blade is advanced over the epiglottis to the tracheal entrance then lifted in order to view the vocal cords.

      The RIGHT-SIDED Macintosh blade is used in adults while the left-sided blade may be used in conditions that make intubation with standard blade difficult e.g. facial deformities.

      The McCoy laryngoscope is based on the STANDARD MACINTOSH blade not Robertshaw’s. It has a lever operated hinged tip, which improves the view during laryngoscopy.

      Polio blade is mounted at an angle of 120-135 degrees to the handle. Originally designed for use during the polio epidemic ​in intubation patients within iron lung ventilators, it is now useful in patients with conditions like breast hypertrophy, barrel chest, and restricted neck mobility.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds

SESSION STATS - PERFORMANCE PER SPECIALTY

Anaesthesia Related Apparatus (3/5) 60%
Passmed