00
Correct
00
Incorrect
00 : 00 : 00
Session Time
00 : 00
Average Question Time ( Mins)
  • Question 1 - For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using...

    Correct

    • For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using 100% oxygen and a fresh gas flow equal to the patient's minute ventilation.

      Which would be the most suitable choice of anaesthetic breathing system in this situation?

      Your Answer: Mapleson A system

      Explanation:

      The Mapleson A (Magill) and coaxial version of the Mapleson A system (Lack circuit) are more efficient for spontaneous breathing than any of the other Mapleson circuits. The fresh gas flow (FGF) required to prevent rebreathing is slightly greater than the alveolar minute ventilation (4-5 litres/minute). This is delivered to the patient through the outer coaxial tube and exhaust gases are moved to the scavenging system through the inner tube. In the Lack circuit, the expiratory valve is located close to the common gas outlet away from the patient end. This is the main advantage of the Lack circuit over the Mapleson A circuit.

      The Mapleson E circuit is a modification of the Ayres T piece and the FGF required to prevent rebreathing is 1.5-2 times the patient’s minute volume.

      The Bain circuit is the coaxial version of the Mapleson D circuit.

      The FGF for spontaneous respiration to avoid rebreathing is 160-200 ml/kg/minute.

      The FGF for controlled ventilation to avoid rebreathing is 70-100 ml/kg/min.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      345.2
      Seconds
  • Question 2 - Regarding tracheal tubes, which of the following statements are true? ...

    Incorrect

    • Regarding tracheal tubes, which of the following statements are true?

      Your Answer: Oxford tubes have a thin wall and are narrower for a given internal diameter

      Correct Answer: Uncuffed RAE tubes have two Murphy eyes

      Explanation:

      Tracheal tubes are made of either disposable plastic or reusable red rubber.

      The tube size refers to the internal diameter (ID) in mm which is marked on the outside of the tube (some manufacturers mark the external diameter on the outside).

      Plastic tubes have a radiopaque line spanning the entire length of the tube, which allows their position to be identified on x-rays. The bevel located at the end of the tube is left-facing and oval in shape, which improves the view of the vocal cords during intubation.

      Oxford tubes are L-shaped and have a bevel that faces posteriorly. They have thick walls that increase the external diameter, making for a wider internal diameter.

      RAE (Ring, Adair, and Elwyn) tubes are preformed and can either be north or south facing and cuffed or uncuffed. The cuffed RAE tubes have one Murphy eye, whereas the uncuffed has two Murphy eyes. Uncuffed tubes are primarily used in paediatric anaesthesia and the two Murphy eyes ensure adequate ventilation- should the tube be too long.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      1647
      Seconds
  • Question 3 - The equipment used for patient monitoring in theatre and intensive care settings have...

    Incorrect

    • The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients.
      Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?

      Your Answer: I (BF)

      Correct Answer: II (CF)

      Explanation:

      Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
      The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
      II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
      III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
      Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
      B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
      BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
      CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      8298.1
      Seconds
  • Question 4 - Concerning the physical principles of temperature measurement by a thermocouple, which of the...

    Incorrect

    • Concerning the physical principles of temperature measurement by a thermocouple, which of the following best describes it?

      Your Answer: The semiconductor at the measuring end has a resistance proportional to temperature

      Correct Answer: The bimetallic strip has a junction potential proportional to temperature

      Explanation:

      A thermocouple, or a thermal junction, is temperature measuring device consisting of a pair of dissimilar metal (bimetallic) wires or strips joined together. Typically, copper and constantan (an alloy of 55% copper and 45% nickel) are used. When there is contact between these metals, a small voltage is generated in the order of millivolts. The magnitude of the thermojunction electromotive force (emf) is proportional to applied temperature (the Seebeck effect). This physical principle is applied in the measurement of temperature. The electromotive force at the measuring junction is proportional to temperature.

      Two wires with different coefficients of expansion, joined together, can be used as a switch for thermostatic control.

      Semiconductors are NOT used in thermocouple. The resistance of the measuring junction of a thermocouple is irrelevant.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      75.4
      Seconds
  • Question 5 - Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency.

    For...

    Incorrect

    • Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency.

      For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?

      Your Answer: Bandwidth 0.5-50 Hz, voltage 0.5-10 millivolts

      Correct Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts

      Explanation:

      The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.

      These are the following:

      ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
      EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
      EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.

      Prior to display, these small signals will need to be amplified and processed further.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      74.6
      Seconds
  • Question 6 - A 74-year-old with a VVI pacemaker is undergoing a hip replacement.

    Which of the...

    Incorrect

    • A 74-year-old with a VVI pacemaker is undergoing a hip replacement.

      Which of the following is most likely to predispose him to an electrical hazard?

      Your Answer: Inadvertent contact with an unearthed operating table

      Correct Answer: Use of cutting unipolar diathermy

      Explanation:

      A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).

      EMI has the potential to cause the following: Inhibition of pacing
      Asynchronous pacing
      Reset to backup mode
      Myocardial burns, and
      Trigger VF.

      Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.

      Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.

      Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.

      Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).

      A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      134.8
      Seconds
  • Question 7 - Regarding the carbon dioxide monitoring, which of the following statements is correct? ...

    Correct

    • Regarding the carbon dioxide monitoring, which of the following statements is correct?

      Your Answer: Carbon dioxide absorbs infrared radiation at 4.28 µm

      Explanation:

      Carbon dioxide (CO2), is a carbonic gas made up of two dissimilar atoms, namely one carbon atom and two oxygen atoms. Capnography is a technique used to measure carbon dioxide during a respiratory cycle, and it consists in calculating the concentration of the partial pressure of CO2, through the absorption of the infrared light, namely that CO2 absorbs infrared radiation at a wavelength of 4.28 µm.

      End-tidal CO2 (ETCO2), referring to the level of the carbon dioxide released at the end of an exhaled breath, is required to be continuously monitored, especially in ventilated patients, as it is a sensitive and a non invasive technique that provides immediate information about ventilation, circulation, and metabolism functions. ETCO2 is normally lower than the arterial partial pressure and varies between 0.6 and 0.7 kPa.

      There are two methods used to measure carbon dioxide. The sidestream capnometer method samples gases at a set flow rate (150-200 mL/min) from a sampling area through small diameter tubing, and the mainstream analyser method that uses a direct measurement of the patient exhaled CO2 by a relatively large and heavy sensors. Sidestram method allows the analysis of multiple gases and anaesthetic vapours comparing to the mainstream method that does not allow the measurement of other gases.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      9258
      Seconds
  • Question 8 - You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy...

    Incorrect

    • You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy to rule out appendicitis.

      She has no medical history and does not take any medications on a regular basis. You're going to do a quick sequence induction.

      Which method of preoxygenation is the most effective and efficient?

      Your Answer:

      Correct Answer: Mapleson A circuit with a fresh gas flow of 100 ml/kg

      Explanation:

      Professor Mapleson classified non-rebreathing circuits based on the position of the APL valve, which controls fresh gas flow.

      The Mapleson A (Magill) circuit is most effective in spontaneous breathing, requiring only 70-100 ml/kg (the patient’s minute volume) of fresh gas flow. The patient inhales fresh gas from the reservoir bag and tubing during inspiration. During expiration, the patient adds dead space gas (gas that hasn’t been exchanged) to the tubing and reservoir bag in addition to the fresh gas flow. At the patient’s end, alveolar gas is vented through the APL valve. During the expiratory pause, the fresh gas flow causes more gas to be released.

      The Mapleson A is inefficient during controlled ventilation. Venting occurs during inspiration rather than during the expiratory phase, as it does during spontaneous ventilation. As a result, unless a high fresh gas flow of >20 L/minute is used, alveolar gas is rebreathed.

      During spontaneous ventilation, the Mapleson D circuit is inefficient.

      The oxygen concentration in a Hudson mask is insufficient to allow for adequate pre-oxygenation.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 9 - A strain gauge operates on which of the following principles? ...

    Incorrect

    • A strain gauge operates on which of the following principles?

      Your Answer:

      Correct Answer: Varying resistance of a wire when stretched

      Explanation:

      The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.

      A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.

      Changes in current flow through the resistor are amplified and displayed as a pressure change measure.

      A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 10 - Which of the following statements is true about oxygen face masks? ...

    Incorrect

    • Which of the following statements is true about oxygen face masks?

      Your Answer:

      Correct Answer: The oxygen concentration delivered by high air flow oxygen enrichment devices is not dependent on the respiratory pattern of the patient

      Explanation:

      The normal peak inspiratory flow in healthy individuals is 20-30 L/min during each normal tidal ventilation. This is expected to increase with greater respiratory rate and deeper inspiration.

      Face masks are used to facilitate the delivery of oxygen from a breathing system to a patient. Face masks can be divided into two types: fixed performance or variable performance devices.

      In fixed performance devices (also known as high air flow oxygen enrichment or HAFOE), fixed inspired oxygen concentration is delivered to the patent, independent and greater than that of the patient’s peak inspiratory flow rate (PIFR). No random entrainment is expected to occur at the time of PIFR, hence, the oxygen concentration in HAFOE devices is not dependent on the patient’s respiratory pattern.

      Moreover, in HAFOE masks, the concentration of oxygen at a given oxygen flow rate is determined by the size of the constriction; a device with a greater entrainment aperture delivers a lower oxygen concentration. Therefore, a 40% Venturi device will have lesser entrainment aperture when compared to a 31% Venturi. Venturi masks allow relatively fixed concentrations of supplemental oxygen to be inspired e.g. 24%, 28%, 31%, 35%, 40% and 60% oxygen. These are colour coded and marked with the recommended oxygen flow rate.

      Variable performance devices deliver variable inspired oxygen concentration to the patient, and is dependent on the PIFR. The PIFR can often exceed the flow rate at which oxygen or an oxygen/air mixture is supplied by the device, depending on a patient’s inspiratory effort. In addition, these masks allow expired air to be released through the holes in the sides of the mask. Thus, with increased respiratory rate, rebreathing of alveolar gas from inside the mask may occur.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 11 - International colour coding is used on medical gas cylinders. Other characteristics also play...

    Incorrect

    • International colour coding is used on medical gas cylinders. Other characteristics also play a role in determining the gas's identity within a cylinder.

      Which of the following options best describes a cylinder containing analgesics for obstetrics?

      Your Answer:

      Correct Answer: Blue body, blue/white shoulder, full cylinder; 13700 KPa, gas mixture, requires a dual stage pressure regulator

      Explanation:

      The body of the Entonox cylinder is usually blue (occasionally white), with blue and white shoulders. Entonox contains a 50:50 mixture of oxygen and nitrous oxide, with a full cylinder pressure of 13700 KPa (137 bar). The cylinder is equipped with a two-stage pressure regulator for safe operation.

      The cylinder body and shoulder of nitrous oxide are (French) blue.

      In today’s anaesthetic workstations, carbon dioxide cylinders are no longer used.

      The body of an oxygen cylinder is black, with a white shoulder.

      The white Heliox (21 percent oxygen and 79 percent helium) cylinder has a brown and white shoulder. The administration of this gas mixture, which is less dense than air, is used to reduce turbulence (stridor) of inspiratory flow in patients with upper airway obstruction.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 12 - Dinamap is an automated blood pressure monitoring device. Which of these statements best...

    Incorrect

    • Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?

      Your Answer:

      Correct Answer: The cuff should be positioned at the same level as the heart

      Explanation:

      Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.

      The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.

      The manual BP device is still the gold standard for BP measurement and monitoring.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 13 - Regarding the emergency oxygen flush, which is true? ...

    Incorrect

    • Regarding the emergency oxygen flush, which is true?

      Your Answer:

      Correct Answer: May lead to awareness if used inappropriately

      Explanation:

      When the emergency oxygen flush is pressed, 100% oxygen is supplied from the common gas outlet. This gas bypasses BOTH flowmeters and vaporisers. The flow of oxygen is usually 45 l/min at a PRESSURE OF 400 kPa.

      There is an increased risk of pulmonary barotrauma when the emergency flush is pressed, especially when anaesthetising paediatric patients.

      The inappropriate use of the flush causes dilution of anaesthetic gases and this increases the possibility of anaesthetic awareness .

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 14 - Which of the following is correct for gas pipeline pressure? ...

    Incorrect

    • Which of the following is correct for gas pipeline pressure?

      Your Answer:

      Correct Answer:

      Explanation:

      Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.

      Carbon dioxide and nitric oxide are usually only supplied in cylinders.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 15 - An emergency appendicectomy is being performed on a 20 year old man. For...

    Incorrect

    • An emergency appendicectomy is being performed on a 20 year old man. For maintenance of anaesthesia, he is being ventilated using a circle system with a fresh gas flow (FGF) of 1 L/min (air/oxygen and sevoflurane). The trace on the capnograph shows a normal shape.

      The table below demonstrates the changes in the end-tidal and baseline carbon dioxide measurements of the capnograph at 10 and 20 minutes of anaesthesia maintenance.  
      End-tidal CO2: 4.9 kPa vs 8.4kPa (10 minutes vs 20 minutes)
      Baseline end-tidal CO2: 0.2 kPa vs 2.4kPa

      Pulse 100-107 beats per minute, systolic blood pressure 125-133 mmHg and oxygen saturation 98-99%. 

      Which of the following is the single most important immediate course of action?

      Your Answer:

      Correct Answer: Increase the FGF

      Explanation:

      End-tidal carbon dioxide (ETCO2) monitoring has been an important factor in reducing anaesthesia-related mortality and morbidity. Hypercarbia, or hypercapnia, occurs when levels of CO2 in the blood become abnormally high (Paco2 >45 mm Hg). Hypercarbia is confirmed by arterial blood gas analysis. When using capnography to approximate Paco2, remember that the normal arterial–end-tidal carbon dioxide gradient is roughly 5 mm Hg. Hypercarbia, therefore, occurs when PETco2 is greater than 40 mm Hg.

      The most likely explanation for the changes in capnograph is either exhaustion of the soda lime and a progressive rise in circuit dead space.

      Inspect the soda lime canister for a change in colour of the granules. To overcome soda lime exhaustion, the first step is to increase the fresh gas flow (FGF) (Option A). Then, if need arises, replace the soda lime granules. Other strategies that can work are changing to another circuit or bypassing the soda lime canister, but remember that both these strategies are employed only after increasing FGF first. Exclude other causes of equipment deadspace too.

      There are also other causes for hypercarbia to develop intraoperatively:
      1. Hypoventilation is the most common cause of hypercapnia. A. Inadequate ventilation can occur with spontaneous breathing due to drugs like anaesthetic agents, opioids, residual NMDs, chronic respiratory or neuromuscular disease, cerebrovascular accident.
      B. In controlled ventilation, hypercapnia due to circuit leaks, disconnection or miscalculation of patient’s minute volume.
      2. Rebreathing – Soda lime exhaustion with circle, inadequate fresh gas flow into Mapleson circuits and increased breathing system deadspace.
      3. Endogenous source – Tourniquet release, hypermetabolic states (MH or thyroid storm) and release of vascular clamps.
      4. Exogenous source – Absorption of CO2 from pneumoperitoneum.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 16 - Regarding a paramagnetic oxygen analyser, the following statements are TRUE: ...

    Incorrect

    • Regarding a paramagnetic oxygen analyser, the following statements are TRUE:

      Your Answer:

      Correct Answer: Utilises null deflection

      Explanation:

      The electrons in the outer shell of an oxygen molecule are unpaired, thus it has paramagnetic properties and is attracted into a magnetic field.

      It utilizes null deflection -True
      Null deflection is a crucial principle in paramagnetic analysers (reflected beam of light on two photocells) which gives very accurate results (typically 0.1%).

      It can be used to measure the concentration of diamagnetic gases – False
      Since most other gases are weakly diamagnetic they are repelled by a magnetic field (nitric oxide is also paramagnetic).

      Can measure gases dissolved in the blood – False
      For accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. Therefore, they are unsuitable to measure gases dissolved in blood.

      Does not require calibration – False
      As with most measurement instruments paramagnetic analysers must be calibrated before use.

      E) The readings are unaffected by water vapour – False
      Water vapour affects the readings hence for accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. That is why they are unsuitable to measure dissolved blood gases.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 17 - Which is the most appropriate statement describing the function of flowmeters? ...

    Incorrect

    • Which is the most appropriate statement describing the function of flowmeters?

      Your Answer:

      Correct Answer: Constant pressure, variable orifice - Heidbrink flowmeters

      Explanation:

      There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 18 - A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted...

    Incorrect

    • A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted in the surgery department for an elective total thyroidectomy with radical neck dissection. The operation is expected to last for 10 hours.

      Which of the following is the most suitable humidifier to use in an anaesthetic circuit for this case?

      Your Answer:

      Correct Answer: Heat and moisture exchanger (HME)

      Explanation:

      Adequate humidification is vital to maintain homeostasis of the airway. Heat and moisture exchangers conserve some of the exhaled water, heat and return them to inspired gases. Many heat and moisture exchangers also perform bacterial/viral filtration and prevent inhalation of small particles. Heat and moisture exchangers are also called condenser humidifier, artificial nose, etc. Most of them are disposable devices with exchanging medium enclosed in a plastic housing. For adult and paediatric age group different dead space types are available. Heat and moisture exchangers are helpful during anaesthesia and ventilatory breathing system. To reduce the damage of the upper respiratory tract through cooling and dehydration inspiratory air can be heated and humidified, thus preventing the serious complications. Moreover, they are the most appropriate humidification devices used for routine anaesthesia.

      Gases can be bubbled through water to increase humidity. Passing gas through water at room temperature causes the gas to cool due to latent heat of vaporisation. The water bath can be heated. This improves the efficiency of the device and also reduces the incidence of bacterial colonisation.

      Nebulisers use a venturi system which employs the Bernoulli effect. A gas at high flow passes through a constriction causing the gas to accelerate, reducing its potential energy allowing other gases or liquids to be entrained. This can include medications or in the case of humidification, water vapour. The size of the water droplet produced by nebulisation determines where in the airway it is deposited. Standard nebulisers produced droplets of 4 microns in diameter and these are deposited in the upper airway and trachea. Efficacy can be improved by passing the droplets over an anvil which further reduces particle size. The most efficient form of nebuliser is the ultrasonic nebuliser. Here a transducer immersed in water and vibrated at a frequency of 3MHz produces1-2micron droplets. These particles easily reach the bronchioles and provide excellent humidification.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 19 - When compared to unipolar diathermy, which of the following is more specific to...

    Incorrect

    • When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?

      Your Answer:

      Correct Answer: Has a power output of up to 140 joules per second

      Explanation:

      Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).

      The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.

      A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.

      High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).

      The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 20 - Regarding laminar gas flow, which of the following options has the most influence...

    Incorrect

    • Regarding laminar gas flow, which of the following options has the most influence on laminar flow?

      Your Answer:

      Correct Answer: Diameter of tube

      Explanation:

      Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:

      1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
      2. Few directional changes
      3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
      4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
      5. Double flow by doubling pressure as long as the flow pattern remains laminar

      Poiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 21 - Among the following, which statement is true regarding electrical safety in an operation...

    Incorrect

    • Among the following, which statement is true regarding electrical safety in an operation theatre?

      Your Answer:

      Correct Answer: The higher the frequency of the current the less risk to the patient

      Explanation:

      The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.

      Electricity can cause morbidity or mortality by one of the following ways:
      (i) electrocution
      (ii) burns
      (iii) ignition of a flammable material, causing a fire or explosion.

      Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.

      Option A: The higher the frequency, the less effects of electrocution on the body.

      Option B & D: Equipment can be classified in classes and types.
      The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
      The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
      Type B:
      can be class I, II or III but the maximum leakage current must not exceed 100 µA. It is therefore not suitable for direct connection to the heart.
      Type BF
      Similar to type B, but uses an isolated (or floating) circuit.
      Type CF
      Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 ?A is sufficient to cause VF.

      Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
      1 mA – tingling pain
      5 mA – pain
      15 mA – tonic muscular contraction
      50 mA – respiratory muscle paralysis
      75 mA – ventricular fibrillation.

      Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 22 - A laser is a device that stimulates atoms or molecules to emit light...

    Incorrect

    • A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment.

      Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?

      Your Answer:

      Correct Answer: Wearing laser protective goggles

      Explanation:

      Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.

      Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.

      Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.

      Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage. Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams. The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury. Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions. Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
      Using short bursts to reduce energy is also not correct as it would still be harmful to eye.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 23 - Which one of the following statement is true regarding United Kingdom gas cylinders?...

    Incorrect

    • Which one of the following statement is true regarding United Kingdom gas cylinders?

      Your Answer:

      Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred

      Explanation:

      Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.

      At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.

      Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.

      The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 24 - Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?...

    Incorrect

    • Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?

      Your Answer:

      Correct Answer: Lack circuit

      Explanation:

      Among the breathing circuits, the Lack circuit is the most efficient for spontaneous breathing.

      An outer coaxial tube is present to deliver fresh air; exhaust air is routed to an inner tube, which is then delivered to a scavenging system. An expiratory valve is seen at the patient end, which is an advantage over other circuits. Moreover, the Lack circuit prevents rebreathing slightly greater than the alveolar minute ventilation at 4-5 litres per minute.

      The Bain circuit prevents rebreathing at 160-200ml/kg per minute, and is a co-axial version of the Mapleson D circuit.

      The Mapleson E circuit prevent rebreathing at a fresh gas flow (FGF) of approximately twice the patient’s normal minute volume. A modification of this, the Mapleson F, has a reservoir bag at the opposite end for the FGF. This circuit is appropriate for paediatric patients with a body weight less than 20 kg.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 25 - Health workers are at increased risks of anaesthetic exposure. Therefore, The Control of...

    Incorrect

    • Health workers are at increased risks of anaesthetic exposure. Therefore, The Control of Substances Hazardous to Health (COSHH) regulations require measures implemented to assess and control the risks related to this exposure.

      Among the following control measures, which one is most likely to limit potentially harmful exposure to anaesthetic agents within an operating theatre?

      Your Answer:

      Correct Answer: Total intravenous anaesthesia (TIVA)

      Explanation:

      Exposure to anaesthetic hazards is one among the occupational exposures in manipulating toxic agents or inhaling toxic gases during anaesthetic practices.

      Toxic gases mainly nitrous oxide, is one of the most gaseous anaesthetic agents that constitutes an important source of pollution. One of the safe and effective technics used in anaesthesia and reducing the amount of pollution is the Total Intravenous Anaesthesia (TIVA) which consists of using opioids in analgesia and propofol for the induction and the maintenance of anaesthesia. It refers to the administration intravenously of an anaesthetic, sedative, and/or tranquilizer. A less polluting but not the best way to get rid of the toxic aesthetic agents is the scavenger system that collects and expels the gas outside the medical environment. Yet, this technique still represents a hazard for the environment and still increase the risk of exposure for the health workers and clinical staff.

      Fume cupboards are also not recommended to use because of their high pollution potency, mainly of the air resulting in a great harm for medical workers.

      Supraglottic airways as well as the Air Changes per Hour technics could be harmful for both patients and health workers, increasing the risks of transmitted diseases, namely nosocomial infections.

      Therefore, the Total Intravenous Anaesthesia technique (TIVA) is most likely to be safe and recommended to use.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 26 - Prior to rapid sequence induction of anaesthesia, a man with a BMI of...

    Incorrect

    • Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated.

      Which method of pre-oxygenation with a tight-fitting face mask is the most effective?

      Your Answer:

      Correct Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths

      Explanation:

      This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:

      Functional residual capacity has been reduced (FRC)
      Increased closing capacity (CC)
      Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
      Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.

      Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 27 - Which of the following is true about the pulse oximeter? ...

    Incorrect

    • Which of the following is true about the pulse oximeter?

      Your Answer:

      Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation

      Explanation:

      The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.

      The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.

      An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.

      The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 28 - Bacteria and viruses that are smaller than 0.1 ?m in diameter can be...

    Incorrect

    • Bacteria and viruses that are smaller than 0.1 ?m in diameter can be filtered out using heat and moisture exchanger (HME) with a typical pore size 0.2 ?m.

      Choose the most appropriate mechanisms of particle capture for most bacteria and viruses.

      Your Answer:

      Correct Answer: Diffusion

      Explanation:

      Warming, humidifying, and filtering inspired anaesthetic gases is done by heat and moisture exchangers (HME) and breathing system filters. They are made of glass fibres materials and are supported by a sturdy frame. Pleating increases the surface area to reduce resistance to air flow and boost efficiency.

      Filters’ effectiveness is determined by the amount and size of particles they keep out of the patient’s airway. The efficiency of filters might be classified as 95, 99.95, or 99.97 percent. Pores with a diameter of 0.2 µm are common. The following are examples of typical particle sizes:
      Red blood cell – 5 µm
      Lymphocyte – 5-8 µm
      Viruses – 0.02-0.3 µm
      Bacteria – 0.5-1 µm
      Depending on particle size, gas flow speed, and charge, particles are collected via a number of processes. Mechanical sieve, interception, diffusion, electrostatic filtration, and inertial impaction are some of the options:

      Sieve:
      The diameter of the particle the filter is supposed to collect is smaller than the apertures of the filter’s fibres.

      Interception:
      When a particle following a gas streamline approaches a fibre within one radius of itself, it becomes attached and captured.
      Diffusion:

      A particle’s random (Brownian) zig-zag path or motion causes it to collide with a fibre.
      By attracting and capturing a particle from within the gas flow, it generates a lower-concentration patch within the gas flow into which another particle diffuses, only to be captured. At low gas velocities and with smaller particles (0.1µm diameter), this is more common.

      Electrostatic:

      These filters use large diameter fibre media and rely on electrostatic charges to improve fine particle removal effectiveness.

      Impaction due to inertia:

      When a particle is too large to respond fast to abrupt changes in streamline direction near a filter fibre, this happens. Because of its inertia, the particle will continue on its original course and collide with the filter fibre. When high gas velocities and dense fibre packing of the filter media are present, this sort of filtration mechanism is most prevalent.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 29 - Which of the following statements is true regarding the Wrights Respirometer? ...

    Incorrect

    • Which of the following statements is true regarding the Wrights Respirometer?

      Your Answer:

      Correct Answer: Measures the minute volume to within an accuracy of +/- 10%

      Explanation:

      A Wrights Respirometer measures the volume of air exhaled over the course of one minute of normal breathing

      It is unidirectional and measures tidal volume and minute volume of gas flow in one direction. It is placed at the expiratory side (lower pressure than inspiratory side therefore lower chances of gas leaks)

      Slits are arranged such that incoming gas will rotate the vane at a rate of 150 revolutions per litre of flowing gas

      The Wright respirometer tends to over-read at high flow rates and under-read at low flows because of mechanical causes like friction and inertia and the accumulation of water vapour

      The ideal flow for accurate readings is 2 L/min for the respirometer. The respirometer reads the tidal volume and minute volume with a ±5–10% accuracy within the range of 4–24 L/min.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 30 - From the following electromagnetic waves, which one has the shortest wavelength? ...

    Incorrect

    • From the following electromagnetic waves, which one has the shortest wavelength?

      Your Answer:

      Correct Answer: X rays

      Explanation:

      Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.

      Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.

      Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds

SESSION STATS - PERFORMANCE PER SPECIALTY

Anaesthesia Related Apparatus (2/7) 29%
Passmed