An emergency appendicectomy is being performed on a 20 year old man. For maintenance of anaesthesia, he is being ventilated using a circle system with a fresh gas flow (FGF) of 1 L/min (air/oxygen and sevoflurane). The trace on the capnograph shows a normal shape.
The table below demonstrates the changes in the end-tidal and baseline carbon dioxide measurements of the capnograph at 10 and 20 minutes of anaesthesia maintenance.
End-tidal CO2: 4.9 kPa vs 8.4kPa (10 minutes vs 20 minutes)
Baseline end-tidal CO2: 0.2 kPa vs 2.4kPa
Pulse 100-107 beats per minute, systolic blood pressure 125-133 mmHg and oxygen saturation 98-99%.
Which of the following is the single most important immediate course of action?Your Answer: Increase the minute volume
Correct Answer: Increase the FGF
Explanation:End-tidal carbon dioxide (ETCO2) monitoring has been an important factor in reducing anaesthesia-related mortality and morbidity. Hypercarbia, or hypercapnia, occurs when levels of CO2 in the blood become abnormally high (Paco2 >45 mm Hg). Hypercarbia is confirmed by arterial blood gas analysis. When using capnography to approximate Paco2, remember that the normal arterial–end-tidal carbon dioxide gradient is roughly 5 mm Hg. Hypercarbia, therefore, occurs when PETco2 is greater than 40 mm Hg.
The most likely explanation for the changes in capnograph is either exhaustion of the soda lime and a progressive rise in circuit dead space.
Inspect the soda lime canister for a change in colour of the granules. To overcome soda lime exhaustion, the first step is to increase the fresh gas flow (FGF) (Option A). Then, if need arises, replace the soda lime granules. Other strategies that can work are changing to another circuit or bypassing the soda lime canister, but remember that both these strategies are employed only after increasing FGF first. Exclude other causes of equipment deadspace too.
There are also other causes for hypercarbia to develop intraoperatively:
1. Hypoventilation is the most common cause of hypercapnia. A. Inadequate ventilation can occur with spontaneous breathing due to drugs like anaesthetic agents, opioids, residual NMDs, chronic respiratory or neuromuscular disease, cerebrovascular accident.
B. In controlled ventilation, hypercapnia due to circuit leaks, disconnection or miscalculation of patient’s minute volume.
2. Rebreathing – Soda lime exhaustion with circle, inadequate fresh gas flow into Mapleson circuits and increased breathing system deadspace.
3. Endogenous source – Tourniquet release, hypermetabolic states (MH or thyroid storm) and release of vascular clamps.
4. Exogenous source – Absorption of CO2 from pneumoperitoneum. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Incorrect
-
Regarding tracheal tubes, which of the following statements are true?
Your Answer: Oxford tubes have a thin wall and are narrower for a given internal diameter
Correct Answer: Uncuffed RAE tubes have two Murphy eyes
Explanation:Tracheal tubes are made of either disposable plastic or reusable red rubber.
The tube size refers to the internal diameter (ID) in mm which is marked on the outside of the tube (some manufacturers mark the external diameter on the outside).
Plastic tubes have a radiopaque line spanning the entire length of the tube, which allows their position to be identified on x-rays. The bevel located at the end of the tube is left-facing and oval in shape, which improves the view of the vocal cords during intubation.
Oxford tubes are L-shaped and have a bevel that faces posteriorly. They have thick walls that increase the external diameter, making for a wider internal diameter.
RAE (Ring, Adair, and Elwyn) tubes are preformed and can either be north or south facing and cuffed or uncuffed. The cuffed RAE tubes have one Murphy eye, whereas the uncuffed has two Murphy eyes. Uncuffed tubes are primarily used in paediatric anaesthesia and the two Murphy eyes ensure adequate ventilation- should the tube be too long.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Correct
-
Which statement is correct about the Mapleson anaesthetic breathing circuits?
Your Answer: Mapleson A is most efficient for spontaneous ventilation
Explanation:Mapleson breathing system (or circuit) analysed five different arrangements of components of the breathing system:
Mapleson A – It is the most efficient for spontaneous respiration. The flow of fresh gas required is 70-85 ml/kg/min, i.e., approximately 5-6 lit./min fresh gas flow for an average adult.
Mapleson B and C – inefficient for both SV and PPV; requires gas flow of two to three times minute volume (100 ml/kg/min). Not commonly used but category C may be used for emergency resuscitation.
Mapleson D – efficient for PPV at gas flow equivalent to patient’s minute volume; the Bain’s circuit is a coaxial version of the Mapleson D
Mapleson E and F – for paediatric use; requires gas flow at two to three times the patient’s minute volume. The Mapleson F consists of an open-ended reservoir bag (Jackson-Rees modification). -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Incorrect
-
Regarding the Manley MP3 ventilator, which statement is true?
Your Answer: Can efficiently ventilate patients with poor pulmonary compliance
Correct Answer: Is a minute volume divider
Explanation:It’s a minute volume divider – True
The Manley MP3 ventilator is classed as a minute volume divider. The entire fresh gas flow or minute volume is delivered to the patient, having been divided into readily set tidal volumes.Can efficiently ventilate patients with poor pulmonary compliance – False
Ventilating patients with poor pulmonary compliance is not easily achieved, which makes it an unsuitable ventilator for a modern ICU.Can generate tidal volume up to 1500ml – False
It can generate tidal volumes up to 1000 ml (not 1500 ml), and the inflation pressure can be adjusted by sliding a weight along a rail.Functions like a Mapleson A system during spontaneous ventilation – False
The ventilator functions like a Mapleson D breathing system (not Mapleson A) during spontaneous ventilation.Has three sets of bellows – False
The fresh gas flow drives the ventilator which allows rapid detection of gas supply failure. It has two sets of bellows (not three) and three unidirectional valves. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Correct
-
For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using 100% oxygen and a fresh gas flow equal to the patient's minute ventilation.
Which would be the most suitable choice of anaesthetic breathing system in this situation?Your Answer: Mapleson A system
Explanation:The Mapleson A (Magill) and coaxial version of the Mapleson A system (Lack circuit) are more efficient for spontaneous breathing than any of the other Mapleson circuits. The fresh gas flow (FGF) required to prevent rebreathing is slightly greater than the alveolar minute ventilation (4-5 litres/minute). This is delivered to the patient through the outer coaxial tube and exhaust gases are moved to the scavenging system through the inner tube. In the Lack circuit, the expiratory valve is located close to the common gas outlet away from the patient end. This is the main advantage of the Lack circuit over the Mapleson A circuit.
The Mapleson E circuit is a modification of the Ayres T piece and the FGF required to prevent rebreathing is 1.5-2 times the patient’s minute volume.
The Bain circuit is the coaxial version of the Mapleson D circuit.
The FGF for spontaneous respiration to avoid rebreathing is 160-200 ml/kg/minute.
The FGF for controlled ventilation to avoid rebreathing is 70-100 ml/kg/min.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 6
Correct
-
Among the following, which statement is true regarding electrical safety in an operation theatre?
Your Answer: The higher the frequency of the current the less risk to the patient
Explanation:The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.
Electricity can cause morbidity or mortality by one of the following ways:
(i) electrocution
(ii) burns
(iii) ignition of a flammable material, causing a fire or explosion.Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.
Option A: The higher the frequency, the less effects of electrocution on the body.
Option B & D: Equipment can be classified in classes and types.
The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
Type B:
can be class I, II or III but the maximum leakage current must not exceed 100 µA. It is therefore not suitable for direct connection to the heart.
Type BF
Similar to type B, but uses an isolated (or floating) circuit.
Type CF
Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 ?A is sufficient to cause VF.Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
1 mA – tingling pain
5 mA – pain
15 mA – tonic muscular contraction
50 mA – respiratory muscle paralysis
75 mA – ventricular fibrillation.Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Correct
-
Which is the most appropriate statement describing the function of flowmeters?
Your Answer: Constant pressure, variable orifice - Heidbrink flowmeters
Explanation:There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Incorrect
-
Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?
Your Answer: Utilises a double cuff oscillotonometric system
Correct Answer: The cuff should be positioned at the same level as the heart
Explanation:Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.
The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.
The manual BP device is still the gold standard for BP measurement and monitoring.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 9
Incorrect
-
What feature is found in the flowmeters of modern anaesthetic machines ?
Your Answer: Have an accuracy of about +/- 4%
Correct Answer: The use of stannic oxide increases the accuracy of flowmeters
Explanation:Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.
Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.
Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.
Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Correct
-
A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment.
Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?Your Answer: Wearing laser protective goggles
Explanation:Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.
Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.
Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.
Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage. Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams. The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury. Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions. Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
Using short bursts to reduce energy is also not correct as it would still be harmful to eye. -
This question is part of the following fields: