  • Question 1 - Many of the processes we deal with in anaesthesia entail a relationship between...

    Incorrect

    • Many of the processes we deal with in anaesthesia entail a relationship between two or more variables.

      Which of the following relationships is a linear relationship?

      Your Answer: The relationship between a decrease in plasma concentration of an intravenous bolus of a drug with time for a single compartment model

      Correct Answer: The relationship between the junction potential and temperature in a thermocouple

      Explanation:

      Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).

      All of the other connections are non-linear.

      For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.

      A sine wave is the relationship between current and degrees or time from a mains power source.

      A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.

      The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      123.7
      Seconds
  • Question 2 - Which one of the following statement is true regarding United Kingdom gas cylinders?...

    Correct

    • Which one of the following statement is true regarding United Kingdom gas cylinders?

      Your Answer: Tensile tests are performed on sections of one cylinder in every hundred

      Explanation:

      Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.

      At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.

      Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.

      The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      18.3
      Seconds
  • Question 3 - A measuring system's response to change is complex, yet it can be mathematically modelled.

    Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?

    ...

    Incorrect

    • A measuring system's response to change is complex, yet it can be mathematically modelled.

      Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?

      Your Answer:

      Correct Answer: Dynamic second-order response

      Explanation:

      The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
      The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:

      Zero-order:
      Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.

      First-order:
      Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.

      Second-order
      Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.

      Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 4 - From the following electromagnetic waves, which one has the shortest wavelength? ...

    Incorrect

    • From the following electromagnetic waves, which one has the shortest wavelength?

      Your Answer:

      Correct Answer: X rays

      Explanation:

      Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.

      Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.

      Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 5 - A laser is a device that stimulates atoms or molecules to emit light...

    Incorrect

    • A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment.

      Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?

      Your Answer:

      Correct Answer: Wearing laser protective goggles

      Explanation:

      Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.

      Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.

      Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.

      Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage. Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams. The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury. Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions. Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
      Using short bursts to reduce energy is also not correct as it would still be harmful to eye.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 6 - With respect to the peripheral nerve stimulators, which one is used to perform...

    Incorrect

    • With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?

      Your Answer:

      Correct Answer:

      Explanation:

      The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.

      There are just 2 leads (not 3); one for the skin and other for the needle.

      Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.

      If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.

      Insulated needles have improved the block success rate, as the current is only conducting through needle tip.

      Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 7 - When compared to unipolar diathermy, which of the following is more specific to...

    Incorrect

    • When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?

      Your Answer:

      Correct Answer: Has a power output of up to 140 joules per second

      Explanation:

      Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).

      The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.

      A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.

      High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).

      The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 8 - Which statement is correct about the Mapleson anaesthetic breathing circuits? ...

    Incorrect

    • Which statement is correct about the Mapleson anaesthetic breathing circuits?

      Your Answer:

      Correct Answer: Mapleson A is most efficient for spontaneous ventilation

      Explanation:

      Mapleson breathing system (or circuit) analysed five different arrangements of components of the breathing system:
      Mapleson A – It is the most efficient for spontaneous respiration. The flow of fresh gas required is 70-85 ml/kg/min, i.e., approximately 5-6 lit./min fresh gas flow for an average adult.
      Mapleson B and C – inefficient for both SV and PPV; requires gas flow of two to three times minute volume (100 ml/kg/min). Not commonly used but category C may be used for emergency resuscitation.
      Mapleson D – efficient for PPV at gas flow equivalent to patient’s minute volume; the Bain’s circuit is a coaxial version of the Mapleson D
      Mapleson E and F – for paediatric use; requires gas flow at two to three times the patient’s minute volume. The Mapleson F consists of an open-ended reservoir bag (Jackson-Rees modification).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 9 - Which of the following is true regarding a laryngoscope? ...

    Incorrect

    • Which of the following is true regarding a laryngoscope?

      Your Answer:

      Correct Answer: The Wisconsin and Seward are examples of straight blade laryngoscopes

      Explanation:

      Direct laryngoscopy are performed using laryngoscopes and they can be classed according to the shape of the blade as curved or straight.

      Miller, Soper, Wisconsin and Seward are examples of straight blade laryngoscopes. Straight blades are commonly used for intubating neonates and infants but can be used in adults too.

      The tip of the miller blade is advanced over the epiglottis to the tracheal entrance then lifted in order to view the vocal cords.

      The RIGHT-SIDED Macintosh blade is used in adults while the left-sided blade may be used in conditions that make intubation with standard blade difficult e.g. facial deformities.

      The McCoy laryngoscope is based on the STANDARD MACINTOSH blade not Robertshaw’s. It has a lever operated hinged tip, which improves the view during laryngoscopy.

      Polio blade is mounted at an angle of 120-135 degrees to the handle. Originally designed for use during the polio epidemic ​in intubation patients within iron lung ventilators, it is now useful in patients with conditions like breast hypertrophy, barrel chest, and restricted neck mobility.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 10 - Concerning the physical principles of temperature measurement by a thermocouple, which of the...

    Incorrect

    • Concerning the physical principles of temperature measurement by a thermocouple, which of the following best describes it?

      Your Answer:

      Correct Answer: The bimetallic strip has a junction potential proportional to temperature

      Explanation:

      A thermocouple, or a thermal junction, is temperature measuring device consisting of a pair of dissimilar metal (bimetallic) wires or strips joined together. Typically, copper and constantan (an alloy of 55% copper and 45% nickel) are used. When there is contact between these metals, a small voltage is generated in the order of millivolts. The magnitude of the thermojunction electromotive force (emf) is proportional to applied temperature (the Seebeck effect). This physical principle is applied in the measurement of temperature. The electromotive force at the measuring junction is proportional to temperature.

      Two wires with different coefficients of expansion, joined together, can be used as a switch for thermostatic control.

      Semiconductors are NOT used in thermocouple. The resistance of the measuring junction of a thermocouple is irrelevant.