-
Question 1
Correct
-
Which of the following statements is TRUE regarding an epidural set?
Your Answer: 19G Tuohy needles have 0.5 cm markings
Explanation:A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings
18G Tuohy catheters are generally 9 to 10cm to hub
Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura
Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure
Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings)
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Correct
-
A strain gauge operates on which of the following principles?
Your Answer: Varying resistance of a wire when stretched
Explanation:The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.
A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.
Changes in current flow through the resistor are amplified and displayed as a pressure change measure.
A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Incorrect
-
Which of the following is correct for gas pipeline pressure?
Your Answer: 4 bar for oxygen
Correct Answer:
Explanation:Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.
Carbon dioxide and nitric oxide are usually only supplied in cylinders.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Incorrect
-
With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?
Your Answer: Prior to injecting the local anaesthetic the ideal current is 1-2 mili amperes at a frequency of 1-2 Hz
Correct Answer:
Explanation:The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.
There are just 2 leads (not 3); one for the skin and other for the needle.
Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.
If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.
Insulated needles have improved the block success rate, as the current is only conducting through needle tip.
Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Incorrect
-
Which one of the following statement is true regarding United Kingdom gas cylinders?
Your Answer: They are manufactured from cast iron
Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred
Explanation:Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.
At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.
Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.
The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 6
Correct
-
Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency.
For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?Your Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts
Explanation:The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.
These are the following:
ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.Prior to display, these small signals will need to be amplified and processed further.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Incorrect
-
What feature is found in the flowmeters of modern anaesthetic machines ?
Your Answer: Oxygen is the first gas to be added to the mixture delivered to the back bar
Correct Answer: The use of stannic oxide increases the accuracy of flowmeters
Explanation:Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.
Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.
Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.
Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Incorrect
-
Regarding the classification of breathing systems, which of the following is true?
Your Answer: The Bain is a coaxial version of the Magill
Correct Answer: The Conway classification describes a functional classification based on whether a CO2 absorber is required
Explanation:Breathing system is an assembly of components which connects patient’s airway to anaesthesia machine through which controlled composition of gas mixture is dispensed. It delivers gas to the patient, removes expired gas and controls the temperature and humidity of the inspired mixture. It allows spontaneous, controlled, or assisted respiration. It may also provide ports for gas sampling, airway pressure, flow and volume monitoring.
Breathing systems have been classified by Conway and Mapleson.
Conway suggested a functional classification:
– Circuits requiring a CO2 absorber
– Circuits not requiring a CO2 absorberWilliam Mapleson designated varying arrangements of breathing system components (masks, breathing tubes, fresh gas flow inlets, adjustable pressure-limiting valves, and reservoir bags) as Mapleson A-E circuits.
Mapleson A: Arranged as FGF inlet, reservoir bag, APL valve, mask.
In this circuit, because the reservoir bag is between the FGF inlet valve and the APL valve, expired gas from the patient may re-enter the system and fill the reservoir bag during controlled ventilation. This is the most efficient system for spontaneous breathing as the FGF must only be equal to a patient’s minute ventilation to prevent rebreathing.Mapleson B: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the FGF inlet is closer to the APL valve, which helps prevent the rebreathing concern in the Mapleson A circuit as above during controlled ventilation.Mapleson C: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the arrangement is the same as the Mapleson B circuit. However, this circuit is shorter as it does not contain elongated corrugated tubing. This circuit also has the FGF inlet close to the APL valve to aid in preventing rebreathing.Mapleson D: Arranged as reservoir bag, APL valve, FGF inlet, and mask.
In this circuit, the arrangement interchanges the FGF inlet and APL valve of the Mapleson A circuit. This system prevents rebreathing by directing FGF towards the APL valve rather than towards the patient during exhalation.Mapleson E: Arranged as corrugated tubing, FGF inlet, and mask.
In this circuit, there is no reservoir bag and no APL valve. Given the inability to alter the pressure of the circuit, this is ideal for spontaneously ventilating neonates or paediatric patients where low-pressure ventilation is desired. The system prevents rebreathing, similar to the Mapleson D circuit.Jackson Rees later modified the Mapleson E by adding an open ended bag, which has since become known as the Mapleson F.
Mapleson F: Arranged as APL valve directly connected to reservoir bag, corrugated tubing, FGF inlet, and mask.
The system prevents rebreathing similarly to Mapleson D by directing FGF towards the APL valve. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 9
Incorrect
-
Many of the processes we deal with in anaesthesia entail a relationship between two or more variables.
Which of the following relationships is a linear relationship?Your Answer: The relationship between the pressure of a fixed mass of gas and its volume at a fixed temperature
Correct Answer: The relationship between the junction potential and temperature in a thermocouple
Explanation:Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).
All of the other connections are non-linear.
For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.
A sine wave is the relationship between current and degrees or time from a mains power source.
A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.
The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Correct
-
A 74-year-old with a VVI pacemaker is undergoing a hip replacement.
Which of the following is most likely to predispose him to an electrical hazard?Your Answer: Use of cutting unipolar diathermy
Explanation:A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).
EMI has the potential to cause the following: Inhibition of pacing
Asynchronous pacing
Reset to backup mode
Myocardial burns, and
Trigger VF.Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.
Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.
Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.
Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).
A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
00
Correct
00
Incorrect
00
:
00
:
00
Session Time
00
:
00
Average Question Time (
Secs)