  • Question 1 - Bacteria and viruses that are smaller than 0.1 ?m in diameter can be...

    Incorrect

    • Bacteria and viruses that are smaller than 0.1 ?m in diameter can be filtered out using heat and moisture exchanger (HME) with a typical pore size 0.2 ?m.

      Choose the most appropriate mechanisms of particle capture for most bacteria and viruses.

      Your Answer: Sieve

      Correct Answer: Diffusion

      Explanation:

      Warming, humidifying, and filtering inspired anaesthetic gases is done by heat and moisture exchangers (HME) and breathing system filters. They are made of glass fibres materials and are supported by a sturdy frame. Pleating increases the surface area to reduce resistance to air flow and boost efficiency.

      Filters’ effectiveness is determined by the amount and size of particles they keep out of the patient’s airway. The efficiency of filters might be classified as 95, 99.95, or 99.97 percent. Pores with a diameter of 0.2 µm are common. The following are examples of typical particle sizes:
      Red blood cell – 5 µm
      Lymphocyte – 5-8 µm
      Viruses – 0.02-0.3 µm
      Bacteria – 0.5-1 µm
      Depending on particle size, gas flow speed, and charge, particles are collected via a number of processes. Mechanical sieve, interception, diffusion, electrostatic filtration, and inertial impaction are some of the options:

      Sieve:
      The diameter of the particle the filter is supposed to collect is smaller than the apertures of the filter’s fibres.

      Interception:
      When a particle following a gas streamline approaches a fibre within one radius of itself, it becomes attached and captured.
      Diffusion:

      A particle’s random (Brownian) zig-zag path or motion causes it to collide with a fibre.
      By attracting and capturing a particle from within the gas flow, it generates a lower-concentration patch within the gas flow into which another particle diffuses, only to be captured. At low gas velocities and with smaller particles (0.1µm diameter), this is more common.

      Electrostatic:

      These filters use large diameter fibre media and rely on electrostatic charges to improve fine particle removal effectiveness.

      Impaction due to inertia:

      When a particle is too large to respond fast to abrupt changes in streamline direction near a filter fibre, this happens. Because of its inertia, the particle will continue on its original course and collide with the filter fibre. When high gas velocities and dense fibre packing of the filter media are present, this sort of filtration mechanism is most prevalent.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      38.5
      Seconds
  • Question 2 - Which medical gas cylinders have the correct colour codes? ...

    Correct

    • Which medical gas cylinders have the correct colour codes?

      Your Answer: Oxygen cylinders have a black body with white shoulders

      Explanation:

      The following are the colour codes for medical gas cylinders:

      Oxygen cylinder has a dark body with white shoulders.

      Nitrous oxide is French blue. Air encompasses a grey body with dark and white quarters on the shoulders.

      Entonox contains a French blue body with white and blue quarters on the shoulders.

      Carbon dioxide barrels are grey in colour.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      51.3
      Seconds
  • Question 3 - Concerning the physical principles of temperature measurement by a thermocouple, which of the...

    Correct

    • Concerning the physical principles of temperature measurement by a thermocouple, which of the following best describes it?

      Your Answer: The bimetallic strip has a junction potential proportional to temperature

      Explanation:

      A thermocouple, or a thermal junction, is temperature measuring device consisting of a pair of dissimilar metal (bimetallic) wires or strips joined together. Typically, copper and constantan (an alloy of 55% copper and 45% nickel) are used. When there is contact between these metals, a small voltage is generated in the order of millivolts. The magnitude of the thermojunction electromotive force (emf) is proportional to applied temperature (the Seebeck effect). This physical principle is applied in the measurement of temperature. The electromotive force at the measuring junction is proportional to temperature.

      Two wires with different coefficients of expansion, joined together, can be used as a switch for thermostatic control.

      Semiconductors are NOT used in thermocouple. The resistance of the measuring junction of a thermocouple is irrelevant.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      139.5
      Seconds
  • Question 4 - A 240 volt alternating current (AC) socket from a wall is used to...

    Correct

    • A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator.

      Name the electrical component that converts AC to DC.

      Your Answer: Rectifier

      Explanation:

      There are two types of defibrillators
      AC defibrillator
      DC defibrillator

      AC defibrillator,
      consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.

      DC defibrillator,
      consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.

      Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.

      Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.

      Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.

      In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      89.1
      Seconds
  • Question 5 - When compared to unipolar diathermy, which of the following is more specific to...

    Incorrect

    • When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?

      Your Answer: Has a frequency of 500 KHz - 1 MHz

      Correct Answer: Has a power output of up to 140 joules per second

      Explanation:

      Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).

      The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.

      A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.

      High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).

      The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      83.9
      Seconds
  • Question 6 - Regarding laminar gas flow, which of the following options has the most influence...

    Correct

    • Regarding laminar gas flow, which of the following options has the most influence on laminar flow?

      Your Answer: Diameter of tube

      Explanation:

      Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:

      1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
      2. Few directional changes
      3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
      4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
      5. Double flow by doubling pressure as long as the flow pattern remains laminar

      Poiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      13.2
      Seconds
  • Question 7 - Many of the processes we deal with in anaesthesia entail a relationship between...

    Incorrect

    • Many of the processes we deal with in anaesthesia entail a relationship between two or more variables.

      Which of the following relationships is a linear relationship?

      Your Answer: The relationship between the pressure of a fixed mass of gas and its volume at a fixed temperature

      Correct Answer: The relationship between the junction potential and temperature in a thermocouple

      Explanation:

      Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).

      All of the other connections are non-linear.

      For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.

      A sine wave is the relationship between current and degrees or time from a mains power source.

      A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.

      The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      103.6
      Seconds
  • Question 8 - Which one of the following statement is true regarding United Kingdom gas cylinders?...

    Incorrect

    • Which one of the following statement is true regarding United Kingdom gas cylinders?

      Your Answer: The nominal gauge pressure of full nitrous oxide cylinder at 20°Cis 13700 kPa

      Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred

      Explanation:

      Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.

      At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.

      Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.

      The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      32.4
      Seconds
  • Question 9 - Among the following, which statement is true regarding electrical safety in an operation...

    Correct

    • Among the following, which statement is true regarding electrical safety in an operation theatre?

      Your Answer: The higher the frequency of the current the less risk to the patient

      Explanation:

      The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.

      Electricity can cause morbidity or mortality by one of the following ways:
      (i) electrocution
      (ii) burns
      (iii) ignition of a flammable material, causing a fire or explosion.

      Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.

      Option A: The higher the frequency, the less effects of electrocution on the body.

      Option B & D: Equipment can be classified in classes and types.
      The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
      The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
      Type B:
      can be class I, II or III but the maximum leakage current must not exceed 100 µA. It is therefore not suitable for direct connection to the heart.
      Type BF
      Similar to type B, but uses an isolated (or floating) circuit.
      Type CF
      Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 ?A is sufficient to cause VF.

      Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
      1 mA – tingling pain
      5 mA – pain
      15 mA – tonic muscular contraction
      50 mA – respiratory muscle paralysis
      75 mA – ventricular fibrillation.

      Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      16.7
      Seconds
  • Question 10 - Which of the following is true regarding a laryngoscope? ...

    Incorrect

    • Which of the following is true regarding a laryngoscope?

      Your Answer: A left-sided Macintosh blade in routinely used in adults

      Correct Answer: The Wisconsin and Seward are examples of straight blade laryngoscopes

      Explanation:

      Direct laryngoscopy are performed using laryngoscopes and they can be classed according to the shape of the blade as curved or straight.

      Miller, Soper, Wisconsin and Seward are examples of straight blade laryngoscopes. Straight blades are commonly used for intubating neonates and infants but can be used in adults too.

      The tip of the miller blade is advanced over the epiglottis to the tracheal entrance then lifted in order to view the vocal cords.

      The RIGHT-SIDED Macintosh blade is used in adults while the left-sided blade may be used in conditions that make intubation with standard blade difficult e.g. facial deformities.

      The McCoy laryngoscope is based on the STANDARD MACINTOSH blade not Robertshaw’s. It has a lever operated hinged tip, which improves the view during laryngoscopy.

      Polio blade is mounted at an angle of 120-135 degrees to the handle. Originally designed for use during the polio epidemic ​in intubation patients within iron lung ventilators, it is now useful in patients with conditions like breast hypertrophy, barrel chest, and restricted neck mobility.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      45.5
      Seconds