Question 1
Incorrect
-
What feature is found in the flowmeters of modern anaesthetic machines ?
Your Answer: Oxygen is the first gas to be added to the mixture delivered to the back bar
Correct Answer: The use of stannic oxide increases the accuracy of flowmeters
Explanation:Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.
Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.
Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.
Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Incorrect
-
With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?
Your Answer:
Correct Answer:
Explanation:The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.
There are just 2 leads (not 3); one for the skin and other for the needle.
Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.
If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.
Insulated needles have improved the block success rate, as the current is only conducting through needle tip.
Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Incorrect
-
When the volume of gas is measured, its value should be supported by a suitable abbreviation. From the following options, what does the STPD stand for?
Your Answer:
Correct Answer: Standard Temperature and Pressure, Dry
Explanation:Gas is composed of large numbers of molecules moving in random directions, separated by distances. They undergo perfectly elastic collisions with each other and the walls of a container and transfer kinetic energy in form of heat. These assumptions bring the characteristics of gases within the range and reasonable approximation to a real gas, particularly how any change in temperature and pressure affect the behaviour of gas. According to different theories and laws proposed, mathematical equations are derived to calculate the volume of gas, also different abbreviations are being used according to given conditions. The abbreviations used are ATP, BTPS, and STPD.
ATP stands for ambient temperature and barometric pressure, it is used to describe the conditions under which volume of gas is measured.
BTPS stands for body temperature and pressure saturated with water vapor. These are conditions under which volume of gas exist and all results of lung volume determination should be quoted at BTPS.
STPD stands for standard temperature and pressure, dry (0C and 760 mm Hg). These are the conditions that are used to describe quantities of individual gases exchanged in the lungs. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Incorrect
-
Many of the processes we deal with in anaesthesia entail a relationship between two or more variables.
Which of the following relationships is a linear relationship?Your Answer:
Correct Answer: The relationship between the junction potential and temperature in a thermocouple
Explanation:Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).
All of the other connections are non-linear.
For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.
A sine wave is the relationship between current and degrees or time from a mains power source.
A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.
The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Incorrect
-
A sevoflurane vaporiser with a 2 percent setting and a 200 kPa ambient pressure is used.
At this pressure, which of the following options best represents vaporiser output?Your Answer:
Correct Answer: The output is 1% because the saturated pressure of sevoflurane is unaffected by ambient pressure
Explanation:Ambient pressure has no effect on a volatile agent’s saturated vapour pressure (SVP). At a temperature of 20°C, the SVP of sevoflurane is approximately 21 kPa, or 21% of atmospheric pressure (100 kPa).
The SVP of sevoflurane remains the same when the ambient pressure is doubled to 200 kPa, but the output of the vaporiser is halved, now 21 percent of 200 kPa, equalling 10.5 percent. The vaporiser’s output has increased to 1%, but the partial pressure output has remained unchanged. The splitting ratio will not change because it is determined by temperature changes.
Calculations can be made as follows:
Vaporizer output % (ambient pressure) = % volatile (calibrated) x 100 kPa calibrated pressure/ambient pressure
2% = 2% (dialled) × 100/100
2% of 100 = 2 kPaAltitude, pressure 50 kPa
4% = 2% (dialled) × 100/50
4% of 50 = 2 kPaHigh pressure at 200 kPa
1% = 2% (dialled) × 100/200
1% of 200 = 2 kPaSevoflurane has a boiling point of 58°C and, unlike desflurane (which has a boiling point of 22.8°C), does not need to be heated and pressurised with a Tec 6 vaporiser.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 6
Incorrect
-
All of the following statements about pH electrode are incorrect except:
Your Answer:
Correct Answer: A semi-permeable membrane reduces protein contamination
Explanation:Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.
The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.
Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.
Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Incorrect
-
When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?
Your Answer:
Correct Answer: Has a power output of up to 140 joules per second
Explanation:Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).
The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.
A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.
High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).
The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Incorrect
-
Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated.
Which method of pre-oxygenation with a tight-fitting face mask is the most effective?Your Answer:
Correct Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths
Explanation:This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:
Functional residual capacity has been reduced (FRC)
Increased closing capacity (CC)
Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 9
Incorrect
-
Regarding the classification of breathing systems, which of the following is true?
Your Answer:
Correct Answer: The Conway classification describes a functional classification based on whether a CO2 absorber is required
Explanation:Breathing system is an assembly of components which connects patient’s airway to anaesthesia machine through which controlled composition of gas mixture is dispensed. It delivers gas to the patient, removes expired gas and controls the temperature and humidity of the inspired mixture. It allows spontaneous, controlled, or assisted respiration. It may also provide ports for gas sampling, airway pressure, flow and volume monitoring.
Breathing systems have been classified by Conway and Mapleson.
Conway suggested a functional classification:
– Circuits requiring a CO2 absorber
– Circuits not requiring a CO2 absorberWilliam Mapleson designated varying arrangements of breathing system components (masks, breathing tubes, fresh gas flow inlets, adjustable pressure-limiting valves, and reservoir bags) as Mapleson A-E circuits.
Mapleson A: Arranged as FGF inlet, reservoir bag, APL valve, mask.
In this circuit, because the reservoir bag is between the FGF inlet valve and the APL valve, expired gas from the patient may re-enter the system and fill the reservoir bag during controlled ventilation. This is the most efficient system for spontaneous breathing as the FGF must only be equal to a patient’s minute ventilation to prevent rebreathing.Mapleson B: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the FGF inlet is closer to the APL valve, which helps prevent the rebreathing concern in the Mapleson A circuit as above during controlled ventilation.Mapleson C: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the arrangement is the same as the Mapleson B circuit. However, this circuit is shorter as it does not contain elongated corrugated tubing. This circuit also has the FGF inlet close to the APL valve to aid in preventing rebreathing.Mapleson D: Arranged as reservoir bag, APL valve, FGF inlet, and mask.
In this circuit, the arrangement interchanges the FGF inlet and APL valve of the Mapleson A circuit. This system prevents rebreathing by directing FGF towards the APL valve rather than towards the patient during exhalation.Mapleson E: Arranged as corrugated tubing, FGF inlet, and mask.
In this circuit, there is no reservoir bag and no APL valve. Given the inability to alter the pressure of the circuit, this is ideal for spontaneously ventilating neonates or paediatric patients where low-pressure ventilation is desired. The system prevents rebreathing, similar to the Mapleson D circuit.Jackson Rees later modified the Mapleson E by adding an open ended bag, which has since become known as the Mapleson F.
Mapleson F: Arranged as APL valve directly connected to reservoir bag, corrugated tubing, FGF inlet, and mask.
The system prevents rebreathing similarly to Mapleson D by directing FGF towards the APL valve. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Incorrect
-
Which of the following statements is TRUE regarding an epidural set?
Your Answer:
Correct Answer: 19G Tuohy needles have 0.5 cm markings
Explanation:A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings
18G Tuohy catheters are generally 9 to 10cm to hub
Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura
Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure
Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings)
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 11
Incorrect
-
Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency.
For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?Your Answer:
Correct Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts
Explanation:The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.
These are the following:
ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.Prior to display, these small signals will need to be amplified and processed further.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 12
Incorrect
-
Which of the following is correct for gas pipeline pressure?
Your Answer:
Correct Answer:
Explanation:Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.
Carbon dioxide and nitric oxide are usually only supplied in cylinders.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 13
Incorrect
-
Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?
Your Answer:
Correct Answer: The cuff should be positioned at the same level as the heart
Explanation:Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.
The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.
The manual BP device is still the gold standard for BP measurement and monitoring.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 14
Incorrect
-
From the following statements, which is true of humidification?
Your Answer:
Correct Answer: Ultrasonic humidifier can achieve greater than 100% relative humidity
Explanation:Increasing temperature increases the amount of water vapour contained in air; for example, at 20°C, air contains about 17 g/m3, and at 37°C, air contains about 44 g/m3. The wet and dry bulb hygrometer, like the hair hygrometer, measures relative humidity.
Under normal operating conditions, Heat and moisture exchangers (HMEs) allows relative humidity of up to 70% to be achieved. Mucus can impair their performance, and they should not be used for longer than 24 hours.
Hot water bath humidifiers might cause scalding, condensed water in the tubing can interfere with gas flow, and there is a danger of infection.
The ultrasonic humidifier operates at roughly 2 MHz and may attain relative humidity levels much above 100%.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 15
Incorrect
-
Which one of the following statement is true regarding United Kingdom gas cylinders?
Your Answer:
Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred
Explanation:Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.
At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.
Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.
The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 16
Incorrect
-
A measuring system's response to change is complex, yet it can be mathematically modelled.
Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?
Your Answer:
Correct Answer: Dynamic second-order response
Explanation:The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:Zero-order:
Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.First-order:
Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.Second-order
Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 17
Incorrect
-
Which of the following is true about the pulse oximeter?
Your Answer:
Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation
Explanation:The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.
The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.
An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.
The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 18
Incorrect
-
Which of the following statements is true regarding the Wrights Respirometer?
Your Answer:
Correct Answer: Measures the minute volume to within an accuracy of +/- 10%
Explanation:A Wrights Respirometer measures the volume of air exhaled over the course of one minute of normal breathing
It is unidirectional and measures tidal volume and minute volume of gas flow in one direction. It is placed at the expiratory side (lower pressure than inspiratory side therefore lower chances of gas leaks)
Slits are arranged such that incoming gas will rotate the vane at a rate of 150 revolutions per litre of flowing gas
The Wright respirometer tends to over-read at high flow rates and under-read at low flows because of mechanical causes like friction and inertia and the accumulation of water vapour
The ideal flow for accurate readings is 2 L/min for the respirometer. The respirometer reads the tidal volume and minute volume with a ±5–10% accuracy within the range of 4–24 L/min.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 19
Incorrect
-
You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy to rule out appendicitis.
She has no medical history and does not take any medications on a regular basis. You're going to do a quick sequence induction.
Which method of preoxygenation is the most effective and efficient?Your Answer:
Correct Answer: Mapleson A circuit with a fresh gas flow of 100 ml/kg
Explanation:Professor Mapleson classified non-rebreathing circuits based on the position of the APL valve, which controls fresh gas flow.
The Mapleson A (Magill) circuit is most effective in spontaneous breathing, requiring only 70-100 ml/kg (the patient’s minute volume) of fresh gas flow. The patient inhales fresh gas from the reservoir bag and tubing during inspiration. During expiration, the patient adds dead space gas (gas that hasn’t been exchanged) to the tubing and reservoir bag in addition to the fresh gas flow. At the patient’s end, alveolar gas is vented through the APL valve. During the expiratory pause, the fresh gas flow causes more gas to be released.
The Mapleson A is inefficient during controlled ventilation. Venting occurs during inspiration rather than during the expiratory phase, as it does during spontaneous ventilation. As a result, unless a high fresh gas flow of >20 L/minute is used, alveolar gas is rebreathed.
During spontaneous ventilation, the Mapleson D circuit is inefficient.
The oxygen concentration in a Hudson mask is insufficient to allow for adequate pre-oxygenation.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 20
Incorrect
-
A 74-year-old with a VVI pacemaker is undergoing a hip replacement.
Which of the following is most likely to predispose him to an electrical hazard?Your Answer:
Correct Answer: Use of cutting unipolar diathermy
Explanation:A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).
EMI has the potential to cause the following: Inhibition of pacing
Asynchronous pacing
Reset to backup mode
Myocardial burns, and
Trigger VF.Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.
Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.
Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.
Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).
A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 21
Incorrect
-
Regarding the carbon dioxide monitoring, which of the following statements is correct?
Your Answer:
Correct Answer: Carbon dioxide absorbs infrared radiation at 4.28 µm
Explanation:Carbon dioxide (CO2), is a carbonic gas made up of two dissimilar atoms, namely one carbon atom and two oxygen atoms. Capnography is a technique used to measure carbon dioxide during a respiratory cycle, and it consists in calculating the concentration of the partial pressure of CO2, through the absorption of the infrared light, namely that CO2 absorbs infrared radiation at a wavelength of 4.28 µm.
End-tidal CO2 (ETCO2), referring to the level of the carbon dioxide released at the end of an exhaled breath, is required to be continuously monitored, especially in ventilated patients, as it is a sensitive and a non invasive technique that provides immediate information about ventilation, circulation, and metabolism functions. ETCO2 is normally lower than the arterial partial pressure and varies between 0.6 and 0.7 kPa.
There are two methods used to measure carbon dioxide. The sidestream capnometer method samples gases at a set flow rate (150-200 mL/min) from a sampling area through small diameter tubing, and the mainstream analyser method that uses a direct measurement of the patient exhaled CO2 by a relatively large and heavy sensors. Sidestram method allows the analysis of multiple gases and anaesthetic vapours comparing to the mainstream method that does not allow the measurement of other gases.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 22
Incorrect
-
Which of the following is true regarding a laryngoscope?
Your Answer:
Correct Answer: The Wisconsin and Seward are examples of straight blade laryngoscopes
Explanation:Direct laryngoscopy are performed using laryngoscopes and they can be classed according to the shape of the blade as curved or straight.
Miller, Soper, Wisconsin and Seward are examples of straight blade laryngoscopes. Straight blades are commonly used for intubating neonates and infants but can be used in adults too.
The tip of the miller blade is advanced over the epiglottis to the tracheal entrance then lifted in order to view the vocal cords.
The RIGHT-SIDED Macintosh blade is used in adults while the left-sided blade may be used in conditions that make intubation with standard blade difficult e.g. facial deformities.
The McCoy laryngoscope is based on the STANDARD MACINTOSH blade not Robertshaw’s. It has a lever operated hinged tip, which improves the view during laryngoscopy.
Polio blade is mounted at an angle of 120-135 degrees to the handle. Originally designed for use during the polio epidemic in intubation patients within iron lung ventilators, it is now useful in patients with conditions like breast hypertrophy, barrel chest, and restricted neck mobility.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 23
Incorrect
-
For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using 100% oxygen and a fresh gas flow equal to the patient's minute ventilation.
Which would be the most suitable choice of anaesthetic breathing system in this situation?Your Answer:
Correct Answer: Mapleson A system
Explanation:The Mapleson A (Magill) and coaxial version of the Mapleson A system (Lack circuit) are more efficient for spontaneous breathing than any of the other Mapleson circuits. The fresh gas flow (FGF) required to prevent rebreathing is slightly greater than the alveolar minute ventilation (4-5 litres/minute). This is delivered to the patient through the outer coaxial tube and exhaust gases are moved to the scavenging system through the inner tube. In the Lack circuit, the expiratory valve is located close to the common gas outlet away from the patient end. This is the main advantage of the Lack circuit over the Mapleson A circuit.
The Mapleson E circuit is a modification of the Ayres T piece and the FGF required to prevent rebreathing is 1.5-2 times the patient’s minute volume.
The Bain circuit is the coaxial version of the Mapleson D circuit.
The FGF for spontaneous respiration to avoid rebreathing is 160-200 ml/kg/minute.
The FGF for controlled ventilation to avoid rebreathing is 70-100 ml/kg/min.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 24
Incorrect
-
Which medical gas cylinders have the correct colour codes?
Your Answer:
Correct Answer: Oxygen cylinders have a black body with white shoulders
Explanation:The following are the colour codes for medical gas cylinders:
Oxygen cylinder has a dark body with white shoulders.
Nitrous oxide is French blue. Air encompasses a grey body with dark and white quarters on the shoulders.
Entonox contains a French blue body with white and blue quarters on the shoulders.
Carbon dioxide barrels are grey in colour.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 25
Incorrect
-
Which statement is correct about the Mapleson anaesthetic breathing circuits?
Your Answer:
Correct Answer: Mapleson A is most efficient for spontaneous ventilation
Explanation:Mapleson breathing system (or circuit) analysed five different arrangements of components of the breathing system:
Mapleson A – It is the most efficient for spontaneous respiration. The flow of fresh gas required is 70-85 ml/kg/min, i.e., approximately 5-6 lit./min fresh gas flow for an average adult.
Mapleson B and C – inefficient for both SV and PPV; requires gas flow of two to three times minute volume (100 ml/kg/min). Not commonly used but category C may be used for emergency resuscitation.
Mapleson D – efficient for PPV at gas flow equivalent to patient’s minute volume; the Bain’s circuit is a coaxial version of the Mapleson D
Mapleson E and F – for paediatric use; requires gas flow at two to three times the patient’s minute volume. The Mapleson F consists of an open-ended reservoir bag (Jackson-Rees modification). -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 26
Incorrect
-
A transport ventilator is powered by an air/oxygen mix using a full oxygen cylinder (class CD) with an internal capacity of 2 litres, and pressure of 23,000 kPa, with a gas flow of 4 litres/minute.
The ventilator also has a control resulting in an additional gas consumption of 1 litre/minute.
How long will it take for the cylinder to empty?Your Answer:
Correct Answer: 92 minutes
Explanation:The Drager Oxylog® 1000 is a pneumatically powered, time-dependent, volume-titrated emergency ventilator with a pressure limit. It is compatible with CD cylinder oxygen. The CD cylinder is a strong and lightweight cylinder usually composed of aluminium or Kevlar. The internal cylinder volume is 2 litres, and the pressure of a full cylinder is 230 bar. The volume of the full cylinder is determined by applying Boyle’s law: P1 × V1 = P2 × V2
Where:
P1= pressure of a full cylinder (230 bar)
V1= volume of oxygen at that pressure (2 litres)
P2= final pressure (1 bar), and
V2= volume of oxygen in the full cylinder.Substituting values into the equation:
230 × 2 = 1 x V2
V2 = 460 litres. The flow of fresh gas is 4 litres/minute + 1 litre/minute required by the control, making a total of 5 litres/minute. The amount of time it takes for the cylinder to empty would be the total volume of oxygen in the full cylinder divided by the amount of oxygen expelled per minute: 460/5 = 92, meaning it would take 92 minutes for the cylinder to empty. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 27
Incorrect
-
The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients.
Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?Your Answer:
Correct Answer: II (CF)
Explanation:Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 28
Incorrect
-
A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment.
Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?Your Answer:
Correct Answer: Wearing laser protective goggles
Explanation:Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.
Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.
Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.
Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage. Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams. The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury. Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions. Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
Using short bursts to reduce energy is also not correct as it would still be harmful to eye. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 29
Incorrect
-
Which statement is correct concerning breathing systems?
Your Answer:
Correct Answer: The reservoir bag can limit the pressure in the breathing system to about 40 cm of water
Explanation:Mapleson classified breathing systems into A, B, C, D and E. Jackson-Rees subsequently modified the Mapleson E by adding a double-ended bag to the end of the reservoir tubing, creating the Mapleson F. A Mapleson E or T-piece does not have a reservoir bag.
A Mapleson A system is a very efficient system for use during spontaneous ventilation. However, it is not suitable for use with patients less than 25 kg, due to the increased dead space at the distal / patient end. This system can be modified into a Lack system or coaxial Mapleson A, where the fresh gas flows through an outer tube (30 mm) and exhaled gases flow through the inner tube (14 mm).
The adjustable pressure limiting valve (APL) or expiratory valve allows exhaled gas and excess fresh gas to leave the breathing system. It is a one-way, adjustable spring-loaded valve, and gases escape when the pressure in the system exceeds the valve opening pressure. During spontaneous ventilation a pressure of less than 1 cm of water (0.1 kPa) is needed when the valve is in the open position (not 2 cm of H2O).
The reservoir bag is highly compliant and when over inflated, the rubber bag can limit the pressure in the system to about 40 cm of H2O.
This is due to the law of Laplace, which states that the pressure will fall as the radius of the bag increases:
Pressure = 2 x tension/radius.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 30
Incorrect
-
From the following electromagnetic waves, which one has the shortest wavelength?
Your Answer:
Correct Answer: X rays
Explanation:Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.
Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.
Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-