-
Question 1
Incorrect
-
The Control of Substances Hazardous to Health (COSHH) regulations recommend air supply rates to specific environments. Which of the following statements is true?
Your Answer: Operating theatres receive a volume of 0.15 m3 of air per second
Correct Answer: Preparation rooms receive a volume of 0.1 m3 of air per second
Explanation:Control of Substances Hazardous to Health (COSHH) was established by government under the Health and Safety at Work act in 1989. Their employers work on identification and management of those substances that are dangerous to health. The implications for anaesthetists include gas scavenging, equipment contamination and environmental safety. Adequate ventilation is required in areas where anaesthetic gases are present. The minimum air supply that is legally required in each specific area is: Operating theatres: 0.65 m3/second. Anaesthetic rooms: 0.15 m3/s. Preparation rooms: 0.1 m3/s. Recovery rooms need 15 air changes per hour
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Incorrect
-
Which one of the following statement is true regarding United Kingdom gas cylinders?
Your Answer: The nominal gauge pressure of full nitrous oxide cylinder at 20°Cis 13700 kPa
Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred
Explanation:Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.
At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.
Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.
The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Correct
-
A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted in the surgery department for an elective total thyroidectomy with radical neck dissection. The operation is expected to last for 10 hours.
Which of the following is the most suitable humidifier to use in an anaesthetic circuit for this case?Your Answer: Heat and moisture exchanger (HME)
Explanation:Adequate humidification is vital to maintain homeostasis of the airway. Heat and moisture exchangers conserve some of the exhaled water, heat and return them to inspired gases. Many heat and moisture exchangers also perform bacterial/viral filtration and prevent inhalation of small particles. Heat and moisture exchangers are also called condenser humidifier, artificial nose, etc. Most of them are disposable devices with exchanging medium enclosed in a plastic housing. For adult and paediatric age group different dead space types are available. Heat and moisture exchangers are helpful during anaesthesia and ventilatory breathing system. To reduce the damage of the upper respiratory tract through cooling and dehydration inspiratory air can be heated and humidified, thus preventing the serious complications. Moreover, they are the most appropriate humidification devices used for routine anaesthesia.
Gases can be bubbled through water to increase humidity. Passing gas through water at room temperature causes the gas to cool due to latent heat of vaporisation. The water bath can be heated. This improves the efficiency of the device and also reduces the incidence of bacterial colonisation.
Nebulisers use a venturi system which employs the Bernoulli effect. A gas at high flow passes through a constriction causing the gas to accelerate, reducing its potential energy allowing other gases or liquids to be entrained. This can include medications or in the case of humidification, water vapour. The size of the water droplet produced by nebulisation determines where in the airway it is deposited. Standard nebulisers produced droplets of 4 microns in diameter and these are deposited in the upper airway and trachea. Efficacy can be improved by passing the droplets over an anvil which further reduces particle size. The most efficient form of nebuliser is the ultrasonic nebuliser. Here a transducer immersed in water and vibrated at a frequency of 3MHz produces1-2micron droplets. These particles easily reach the bronchioles and provide excellent humidification.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Incorrect
-
Which of the following statements is TRUE regarding an epidural set?
Your Answer: An 18G Tuohy needle is 8 cm in length
Correct Answer: 19G Tuohy needles have 0.5 cm markings
Explanation:A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings
18G Tuohy catheters are generally 9 to 10cm to hub
Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura
Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure
Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings)
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Incorrect
-
The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients.
Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?Your Answer: III
Correct Answer: II (CF)
Explanation:Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 6
Incorrect
-
All of the following statements about pH electrode are incorrect except:
Your Answer: The reference electrode contains a saturated potassium hydroxide solution
Correct Answer: A semi-permeable membrane reduces protein contamination
Explanation:Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.
The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.
Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.
Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Incorrect
-
Regarding tracheal tubes, which of the following statements are true?
Your Answer: The tube size refers to the external diameter which is marked on the outside of the tube in millimetres
Correct Answer: Uncuffed RAE tubes have two Murphy eyes
Explanation:Tracheal tubes are made of either disposable plastic or reusable red rubber.
The tube size refers to the internal diameter (ID) in mm which is marked on the outside of the tube (some manufacturers mark the external diameter on the outside).
Plastic tubes have a radiopaque line spanning the entire length of the tube, which allows their position to be identified on x-rays. The bevel located at the end of the tube is left-facing and oval in shape, which improves the view of the vocal cords during intubation.
Oxford tubes are L-shaped and have a bevel that faces posteriorly. They have thick walls that increase the external diameter, making for a wider internal diameter.
RAE (Ring, Adair, and Elwyn) tubes are preformed and can either be north or south facing and cuffed or uncuffed. The cuffed RAE tubes have one Murphy eye, whereas the uncuffed has two Murphy eyes. Uncuffed tubes are primarily used in paediatric anaesthesia and the two Murphy eyes ensure adequate ventilation- should the tube be too long.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Correct
-
Which statement most accurately describes the action of increasing the gain in ultrasound imaging?
Your Answer: Amplifies the returning signal
Explanation:A higher frequency ultrasound comes with a better resolution of the digital image. Ultrasound with a frequency of 15 MHz is best used in imaging of superficial organs such as the thyroid gland, muscles, tendons and breasts whereas deep organs are better imaged using a lower frequency of 2-7MHz because of its ability for deeper penetration but lower resolution. These low frequency probes are also used to diagnose ascites, pleural effusions or can be used in echocardiography.
The US probe emits and then absorbs a reflected wave. Similar to brightness control, increasing the gain will amplify the return signal which is then attenuated by the tissue. This increases the signal to noise ratio.
A high frame rate, which basically means the number of times an image is updated onto the screen per second, improves the resolution of a moving 3D image which has become more accurate as the computing power has increased.Widening of the image field can be obtained by altering the penetration depth which is obtained by changing the frequency of the US beam
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 9
Incorrect
-
Which is the most appropriate statement describing the function of flowmeters?
Your Answer:
Correct Answer: Constant pressure, variable orifice - Heidbrink flowmeters
Explanation:There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Incorrect
-
From the following electromagnetic waves, which one has the shortest wavelength?
Your Answer:
Correct Answer: X rays
Explanation:Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.
Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.
Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 11
Incorrect
-
You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy to rule out appendicitis.
She has no medical history and does not take any medications on a regular basis. You're going to do a quick sequence induction.
Which method of preoxygenation is the most effective and efficient?Your Answer:
Correct Answer: Mapleson A circuit with a fresh gas flow of 100 ml/kg
Explanation:Professor Mapleson classified non-rebreathing circuits based on the position of the APL valve, which controls fresh gas flow.
The Mapleson A (Magill) circuit is most effective in spontaneous breathing, requiring only 70-100 ml/kg (the patient’s minute volume) of fresh gas flow. The patient inhales fresh gas from the reservoir bag and tubing during inspiration. During expiration, the patient adds dead space gas (gas that hasn’t been exchanged) to the tubing and reservoir bag in addition to the fresh gas flow. At the patient’s end, alveolar gas is vented through the APL valve. During the expiratory pause, the fresh gas flow causes more gas to be released.
The Mapleson A is inefficient during controlled ventilation. Venting occurs during inspiration rather than during the expiratory phase, as it does during spontaneous ventilation. As a result, unless a high fresh gas flow of >20 L/minute is used, alveolar gas is rebreathed.
During spontaneous ventilation, the Mapleson D circuit is inefficient.
The oxygen concentration in a Hudson mask is insufficient to allow for adequate pre-oxygenation.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 12
Incorrect
-
Which of the following is correct for gas pipeline pressure?
Your Answer:
Correct Answer:
Explanation:Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.
Carbon dioxide and nitric oxide are usually only supplied in cylinders.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 13
Incorrect
-
Many of the processes we deal with in anaesthesia entail a relationship between two or more variables.
Which of the following relationships is a linear relationship?Your Answer:
Correct Answer: The relationship between the junction potential and temperature in a thermocouple
Explanation:Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).
All of the other connections are non-linear.
For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.
A sine wave is the relationship between current and degrees or time from a mains power source.
A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.
The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 14
Incorrect
-
Regarding the classification of breathing systems, which of the following is true?
Your Answer:
Correct Answer: The Conway classification describes a functional classification based on whether a CO2 absorber is required
Explanation:Breathing system is an assembly of components which connects patient’s airway to anaesthesia machine through which controlled composition of gas mixture is dispensed. It delivers gas to the patient, removes expired gas and controls the temperature and humidity of the inspired mixture. It allows spontaneous, controlled, or assisted respiration. It may also provide ports for gas sampling, airway pressure, flow and volume monitoring.
Breathing systems have been classified by Conway and Mapleson.
Conway suggested a functional classification:
– Circuits requiring a CO2 absorber
– Circuits not requiring a CO2 absorberWilliam Mapleson designated varying arrangements of breathing system components (masks, breathing tubes, fresh gas flow inlets, adjustable pressure-limiting valves, and reservoir bags) as Mapleson A-E circuits.
Mapleson A: Arranged as FGF inlet, reservoir bag, APL valve, mask.
In this circuit, because the reservoir bag is between the FGF inlet valve and the APL valve, expired gas from the patient may re-enter the system and fill the reservoir bag during controlled ventilation. This is the most efficient system for spontaneous breathing as the FGF must only be equal to a patient’s minute ventilation to prevent rebreathing.Mapleson B: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the FGF inlet is closer to the APL valve, which helps prevent the rebreathing concern in the Mapleson A circuit as above during controlled ventilation.Mapleson C: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the arrangement is the same as the Mapleson B circuit. However, this circuit is shorter as it does not contain elongated corrugated tubing. This circuit also has the FGF inlet close to the APL valve to aid in preventing rebreathing.Mapleson D: Arranged as reservoir bag, APL valve, FGF inlet, and mask.
In this circuit, the arrangement interchanges the FGF inlet and APL valve of the Mapleson A circuit. This system prevents rebreathing by directing FGF towards the APL valve rather than towards the patient during exhalation.Mapleson E: Arranged as corrugated tubing, FGF inlet, and mask.
In this circuit, there is no reservoir bag and no APL valve. Given the inability to alter the pressure of the circuit, this is ideal for spontaneously ventilating neonates or paediatric patients where low-pressure ventilation is desired. The system prevents rebreathing, similar to the Mapleson D circuit.Jackson Rees later modified the Mapleson E by adding an open ended bag, which has since become known as the Mapleson F.
Mapleson F: Arranged as APL valve directly connected to reservoir bag, corrugated tubing, FGF inlet, and mask.
The system prevents rebreathing similarly to Mapleson D by directing FGF towards the APL valve. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 15
Incorrect
-
A measuring system's response to change is complex, yet it can be mathematically modelled.
Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?
Your Answer:
Correct Answer: Dynamic second-order response
Explanation:The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:Zero-order:
Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.First-order:
Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.Second-order
Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 16
Incorrect
-
Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?
Your Answer:
Correct Answer: Lack circuit
Explanation:Among the breathing circuits, the Lack circuit is the most efficient for spontaneous breathing.
An outer coaxial tube is present to deliver fresh air; exhaust air is routed to an inner tube, which is then delivered to a scavenging system. An expiratory valve is seen at the patient end, which is an advantage over other circuits. Moreover, the Lack circuit prevents rebreathing slightly greater than the alveolar minute ventilation at 4-5 litres per minute.
The Bain circuit prevents rebreathing at 160-200ml/kg per minute, and is a co-axial version of the Mapleson D circuit.
The Mapleson E circuit prevent rebreathing at a fresh gas flow (FGF) of approximately twice the patient’s normal minute volume. A modification of this, the Mapleson F, has a reservoir bag at the opposite end for the FGF. This circuit is appropriate for paediatric patients with a body weight less than 20 kg.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 17
Incorrect
-
Which of the following statements is true regarding the Wrights Respirometer?
Your Answer:
Correct Answer: Measures the minute volume to within an accuracy of +/- 10%
Explanation:A Wrights Respirometer measures the volume of air exhaled over the course of one minute of normal breathing
It is unidirectional and measures tidal volume and minute volume of gas flow in one direction. It is placed at the expiratory side (lower pressure than inspiratory side therefore lower chances of gas leaks)
Slits are arranged such that incoming gas will rotate the vane at a rate of 150 revolutions per litre of flowing gas
The Wright respirometer tends to over-read at high flow rates and under-read at low flows because of mechanical causes like friction and inertia and the accumulation of water vapour
The ideal flow for accurate readings is 2 L/min for the respirometer. The respirometer reads the tidal volume and minute volume with a ±5–10% accuracy within the range of 4–24 L/min.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 18
Incorrect
-
A current flows through a simple electric circuit.
Which of the following electrical component configurations has the greatest potential difference?Your Answer:
Correct Answer: Two 5 ohm resistors in series with a passing current of 10 ampere
Explanation: -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 19
Incorrect
-
A sevoflurane vaporiser with a 2 percent setting and a 200 kPa ambient pressure is used.
At this pressure, which of the following options best represents vaporiser output?Your Answer:
Correct Answer: The output is 1% because the saturated pressure of sevoflurane is unaffected by ambient pressure
Explanation:Ambient pressure has no effect on a volatile agent’s saturated vapour pressure (SVP). At a temperature of 20°C, the SVP of sevoflurane is approximately 21 kPa, or 21% of atmospheric pressure (100 kPa).
The SVP of sevoflurane remains the same when the ambient pressure is doubled to 200 kPa, but the output of the vaporiser is halved, now 21 percent of 200 kPa, equalling 10.5 percent. The vaporiser’s output has increased to 1%, but the partial pressure output has remained unchanged. The splitting ratio will not change because it is determined by temperature changes.
Calculations can be made as follows:
Vaporizer output % (ambient pressure) = % volatile (calibrated) x 100 kPa calibrated pressure/ambient pressure
2% = 2% (dialled) × 100/100
2% of 100 = 2 kPaAltitude, pressure 50 kPa
4% = 2% (dialled) × 100/50
4% of 50 = 2 kPaHigh pressure at 200 kPa
1% = 2% (dialled) × 100/200
1% of 200 = 2 kPaSevoflurane has a boiling point of 58°C and, unlike desflurane (which has a boiling point of 22.8°C), does not need to be heated and pressurised with a Tec 6 vaporiser.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 20
Incorrect
-
Which of the following is true about the pulse oximeter?
Your Answer:
Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation
Explanation:The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.
The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.
An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.
The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 21
Incorrect
-
With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?
Your Answer:
Correct Answer:
Explanation:The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.
There are just 2 leads (not 3); one for the skin and other for the needle.
Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.
If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.
Insulated needles have improved the block success rate, as the current is only conducting through needle tip.
Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 22
Incorrect
-
Regarding the Manley MP3 ventilator, which statement is true?
Your Answer:
Correct Answer: Is a minute volume divider
Explanation:It’s a minute volume divider – True
The Manley MP3 ventilator is classed as a minute volume divider. The entire fresh gas flow or minute volume is delivered to the patient, having been divided into readily set tidal volumes.Can efficiently ventilate patients with poor pulmonary compliance – False
Ventilating patients with poor pulmonary compliance is not easily achieved, which makes it an unsuitable ventilator for a modern ICU.Can generate tidal volume up to 1500ml – False
It can generate tidal volumes up to 1000 ml (not 1500 ml), and the inflation pressure can be adjusted by sliding a weight along a rail.Functions like a Mapleson A system during spontaneous ventilation – False
The ventilator functions like a Mapleson D breathing system (not Mapleson A) during spontaneous ventilation.Has three sets of bellows – False
The fresh gas flow drives the ventilator which allows rapid detection of gas supply failure. It has two sets of bellows (not three) and three unidirectional valves. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 23
Incorrect
-
Regarding laminar gas flow, which of the following options has the most influence on laminar flow?
Your Answer:
Correct Answer: Diameter of tube
Explanation:Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:
1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
2. Few directional changes
3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
5. Double flow by doubling pressure as long as the flow pattern remains laminarPoiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 24
Incorrect
-
An emergency appendicectomy is being performed on a 20 year old man. For maintenance of anaesthesia, he is being ventilated using a circle system with a fresh gas flow (FGF) of 1 L/min (air/oxygen and sevoflurane). The trace on the capnograph shows a normal shape.
The table below demonstrates the changes in the end-tidal and baseline carbon dioxide measurements of the capnograph at 10 and 20 minutes of anaesthesia maintenance. Â
End-tidal CO2: 4.9 kPa vs 8.4kPa (10 minutes vs 20 minutes)
Baseline end-tidal CO2: 0.2 kPa vs 2.4kPa
Pulse 100-107 beats per minute, systolic blood pressure 125-133 mmHg and oxygen saturation 98-99%.Â
Which of the following is the single most important immediate course of action?Your Answer:
Correct Answer: Increase the FGF
Explanation:End-tidal carbon dioxide (ETCO2) monitoring has been an important factor in reducing anaesthesia-related mortality and morbidity. Hypercarbia, or hypercapnia, occurs when levels of CO2 in the blood become abnormally high (Paco2 >45 mm Hg). Hypercarbia is confirmed by arterial blood gas analysis. When using capnography to approximate Paco2, remember that the normal arterial–end-tidal carbon dioxide gradient is roughly 5 mm Hg. Hypercarbia, therefore, occurs when PETco2 is greater than 40 mm Hg.
The most likely explanation for the changes in capnograph is either exhaustion of the soda lime and a progressive rise in circuit dead space.
Inspect the soda lime canister for a change in colour of the granules. To overcome soda lime exhaustion, the first step is to increase the fresh gas flow (FGF) (Option A). Then, if need arises, replace the soda lime granules. Other strategies that can work are changing to another circuit or bypassing the soda lime canister, but remember that both these strategies are employed only after increasing FGF first. Exclude other causes of equipment deadspace too.
There are also other causes for hypercarbia to develop intraoperatively:
1. Hypoventilation is the most common cause of hypercapnia. A. Inadequate ventilation can occur with spontaneous breathing due to drugs like anaesthetic agents, opioids, residual NMDs, chronic respiratory or neuromuscular disease, cerebrovascular accident.
B. In controlled ventilation, hypercapnia due to circuit leaks, disconnection or miscalculation of patient’s minute volume.
2. Rebreathing – Soda lime exhaustion with circle, inadequate fresh gas flow into Mapleson circuits and increased breathing system deadspace.
3. Endogenous source – Tourniquet release, hypermetabolic states (MH or thyroid storm) and release of vascular clamps.
4. Exogenous source – Absorption of CO2 from pneumoperitoneum. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 25
Incorrect
-
Which of the following statements is true about oxygen face masks?
Your Answer:
Correct Answer: The oxygen concentration delivered by high air flow oxygen enrichment devices is not dependent on the respiratory pattern of the patient
Explanation:The normal peak inspiratory flow in healthy individuals is 20-30 L/min during each normal tidal ventilation. This is expected to increase with greater respiratory rate and deeper inspiration.
Face masks are used to facilitate the delivery of oxygen from a breathing system to a patient. Face masks can be divided into two types: fixed performance or variable performance devices.
In fixed performance devices (also known as high air flow oxygen enrichment or HAFOE), fixed inspired oxygen concentration is delivered to the patent, independent and greater than that of the patient’s peak inspiratory flow rate (PIFR). No random entrainment is expected to occur at the time of PIFR, hence, the oxygen concentration in HAFOE devices is not dependent on the patient’s respiratory pattern.
Moreover, in HAFOE masks, the concentration of oxygen at a given oxygen flow rate is determined by the size of the constriction; a device with a greater entrainment aperture delivers a lower oxygen concentration. Therefore, a 40% Venturi device will have lesser entrainment aperture when compared to a 31% Venturi. Venturi masks allow relatively fixed concentrations of supplemental oxygen to be inspired e.g. 24%, 28%, 31%, 35%, 40% and 60% oxygen. These are colour coded and marked with the recommended oxygen flow rate.
Variable performance devices deliver variable inspired oxygen concentration to the patient, and is dependent on the PIFR. The PIFR can often exceed the flow rate at which oxygen or an oxygen/air mixture is supplied by the device, depending on a patient’s inspiratory effort. In addition, these masks allow expired air to be released through the holes in the sides of the mask. Thus, with increased respiratory rate, rebreathing of alveolar gas from inside the mask may occur.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 26
Incorrect
-
A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator.
Name the electrical component that converts AC to DC.Your Answer:
Correct Answer: Rectifier
Explanation:There are two types of defibrillators
AC defibrillator
DC defibrillatorAC defibrillator,
consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.DC defibrillator,
consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.
Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.
Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.
In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 27
Incorrect
-
Concerning the physical principles of temperature measurement by a thermocouple, which of the following best describes it?
Your Answer:
Correct Answer: The bimetallic strip has a junction potential proportional to temperature
Explanation:A thermocouple, or a thermal junction, is temperature measuring device consisting of a pair of dissimilar metal (bimetallic) wires or strips joined together. Typically, copper and constantan (an alloy of 55% copper and 45% nickel) are used. When there is contact between these metals, a small voltage is generated in the order of millivolts. The magnitude of the thermojunction electromotive force (emf) is proportional to applied temperature (the Seebeck effect). This physical principle is applied in the measurement of temperature. The electromotive force at the measuring junction is proportional to temperature.
Two wires with different coefficients of expansion, joined together, can be used as a switch for thermostatic control.
Semiconductors are NOT used in thermocouple. The resistance of the measuring junction of a thermocouple is irrelevant.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 28
Incorrect
-
Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated.
Which method of pre-oxygenation with a tight-fitting face mask is the most effective?Your Answer:
Correct Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths
Explanation:This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:
Functional residual capacity has been reduced (FRC)
Increased closing capacity (CC)
Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 29
Incorrect
-
When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?
Your Answer:
Correct Answer: Has a power output of up to 140 joules per second
Explanation:Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).
The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.
A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.
High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).
The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 30
Incorrect
-
A 74-year-old with a VVI pacemaker is undergoing a hip replacement.
Which of the following is most likely to predispose him to an electrical hazard?Your Answer:
Correct Answer: Use of cutting unipolar diathermy
Explanation:A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).
EMI has the potential to cause the following: Inhibition of pacing
Asynchronous pacing
Reset to backup mode
Myocardial burns, and
Trigger VF.Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.
Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.
Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.
Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).
A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
00
Correct
00
Incorrect
00
:
00
:
00
Session Time
00
:
00
Average Question Time (
Secs)