00
Correct
00
Incorrect
00 : 00 : 00
Session Time
00 : 00
Average Question Time ( Mins)
  • Question 1 - A 32-year-old male is admitted to the critical care unit. He has suffered...

    Correct

    • A 32-year-old male is admitted to the critical care unit. He has suffered a heroin overdose and requires intubation and ventilatory support.

      What would be his predicted total static compliance (lung and chest wall) measurements.

      Your Answer: 100 ml/cmH2O

      Explanation:

      Static lung compliance refers to the change in volume within the lung per given change in unit pressure. It is usually measured when air flow is absent, such as during pauses in inhalation and exhalation.

      It is a combination of:

      Chest wall compliance: normal value is 200 mL/cmH2O
      Lung tissue compliance: normal value is 200 mL/ cmH2O

      It is represented mathematically as:

      1/Crs = 1/Cl + 1/Ccw

      Where,

      Crs = total compliance of the respiratory system
      Cl = compliance of the lung
      Ccw = compliance of the chest wall

      Therefore in this case:

      1/Crs = 1/200 + 1/200

      1/Crs = 0.005 + 0.005 = 0.01

      1/Ct = 0.01

      Rearranging equation gives:

      Ct = 1/0.01 = 100 mL/cmH2O.

    • This question is part of the following fields:

      • Clinical Measurement
      93.1
      Seconds
  • Question 2 - The following statements are about capnography. Which of them is true? ...

    Incorrect

    • The following statements are about capnography. Which of them is true?

      Your Answer: Peak infrared absorption of carbon dioxide is at a wave length of 5.28 micrometres

      Correct Answer: Collision broadening is due to presence of other polyatomic molecules

      Explanation:

      Capnography is the non-invasive measurement and pictorial display of inhaled and exhaled carbon dioxide (CO2) partial pressure.

      It is depicted graphically as the concentration of CO2 over time.

      It is used in disease diagnosis, determining disease severity, assessing response to treatment and is the best method to for indicating when an endotracheal tube is placed in the trachea after intubation.

      The wavelength of IR light usually absorbed by nitrous oxide is between 4.4-4.6?m (very close to that of CO2). Its absorption of wavelengths at 3.9 ?m is much weaker. It causes a measurable deficit of 0.1% for every 10% of nitrous oxide. The maximal wavelength of infrared (IR) light absorbed by carbon monoxide is 4.7 ?m. The volatile agents have strong absorption bands at 3.3 ?m and throughout the ranges 8-12 ?m.

      IR light is not absorbed by oxygen (O2), but O2 and CO2 molecules are constantly colliding which interrupts the absorption of IR light by CO2. This increases the band of absorption, that is the Collison or pressure broadening). An oxygen percentage of 95 will result in a 0.5 percentage fall in CO2 measure.

      IR light is also absorbed by water vapour which will result in an overlap of the absorption band, collision broadening and a dilution of partial pressure. This is why water trap and water permeable tubing is recommended for use as it reduces measurement inaccuracies.

      The use of multi-gas analysers of modern gases also help reduce the effects of collision broadening.

      Beer’s law is also applied in this system as an increase in the concentrations of CO2 causes a decrease in the amount of IR able to pass through the gas. This IR light is what generated the signal that is analysed for display.

      The capnograph can indicate oesophageal intubation, but cannot determine if it is endotracheal or endobronchial. For this, auscultation is used.

    • This question is part of the following fields:

      • Clinical Measurement
      30.2
      Seconds
  • Question 3 - A pulmonary function tests results are:

    Measurement Predicted result Test result
    Forced vital capacity (FVC)...

    Correct

    • A pulmonary function tests results are:

      Measurement Predicted result Test result
      Forced vital capacity (FVC) (btps) 3.85 2.36
      Forced expiratory volume in 1 second (FEV1) (btps) 3.34 0.97
      FEV1/FVC ratio % (btps) 85.1 39.9
      Peak expiratory flow (PEF) (L/second) 7.33 2.11
      Maximum voluntary ventilation (MVV) (L/minute) 116 44.4

      What does this indicate?

      Your Answer: Moderate restrictive and severe obstructive picture

      Explanation:

      Severity of a reduction in restrictive defect (%FVC) or obstructive defect (%FEV1/FVC) predicted are classified as follows:

      Mild 70-80%
      Moderate 60-69%
      Moderately severe 50-59%
      Severe 35-49%
      Very severe <35% This patient has a mixed deficit with a severe obstructive deficit as %FEV1/FVC predicted is 46.9% and a moderate restrictive deficit as %FVC of predicted is 61.3 FEV1/FVC ratio 80% < predicted and VC < 80% = mixed picture. FEV1/FVC ratio 80% < predicted and VC > 80% = obstructive picture.

      FEV1/FVC ratio 80% > predicted and VC > 80% = normal picture.

      FEV1/FVC ratio 80% > predicted and VC < 80% predicted= restrictive picture.

    • This question is part of the following fields:

      • Clinical Measurement
      15.2
      Seconds
  • Question 4 - Which measurements is the most accurate for predicting fluid responsiveness? ...

    Correct

    • Which measurements is the most accurate for predicting fluid responsiveness?

      Your Answer: Change in stroke volume following passive straight leg raise

      Explanation:

      The passive leg raising (PLR) manoeuvre is a method of altering left and right ventricular preload and it is done with real-time measurement of stroke volume. It is a simple, quick, relatively unbiased, and accurate bedside test to guide fluid management and avoid fluid overload.

      Pulse pressure variation (PPV), Stroke volume variation (SVV), superior vena cava diameter variation (threshold 36%) and end-expiratory occlusion test are used for dynamic tests of fluid responsiveness.

      PPV is derived peripherally from the arterial pressure waveform.

      Stroke volume variation (SVV) can be derived peripherally through pulse contour analysis of the arterial waveform. PPV and SVV have a threshold of 12% but since they are not used in patients who have cardiac arrhythmias, are spontaneous breathing, and in ventilated patients with low lung compliance and tidal volumes, they are of limited value.

      The tests of fluid responsiveness’ accuracy is determined by calculating the area under the receiver operating characteristic curve (UROC) obtained by plotting the sensitivity of the parameter in predicting fluid responsiveness vs. 1-specificity.

      Under optimal conditions, the ability to determine the need for fluid is best with PPV>SVV>LVEDA>CVP.

      Central venous pressure (CVP) is a static test of preload (not preload responsiveness) and a key determinant of cardiac function. The left ventricular end-diastolic area (LVEDA) a static test of fluid responsiveness, is derived using echocardiography

    • This question is part of the following fields:

      • Clinical Measurement
      85.8
      Seconds
  • Question 5 - A 30 year old male was the victim of an electrocution injury and...

    Incorrect

    • A 30 year old male was the victim of an electrocution injury and has been treated. The option that best describes the current levels for this injury is:

      Your Answer: Respiratory arrest - 75 mA

      Correct Answer: Tonic muscle contraction - 15 mA

      Explanation:

      There are different effects of electrocution and these can be shown in the table below.

      Current Effect
      1 mA Tingling
      5 mA Pain
      15 mA Tonic muscle contraction
      50 mA Respiratory arrest
      100 mA Ventricular fibrillation and cardiac arrest

    • This question is part of the following fields:

      • Clinical Measurement
      34.7
      Seconds
  • Question 6 - The diaphragm is a muscle that is relatively resistant to non-depolarizing neuromuscular blockade's...

    Correct

    • The diaphragm is a muscle that is relatively resistant to non-depolarizing neuromuscular blockade's effects.

      When these muscle relaxants are used, which of the following peripheral nerve stimulator twitch patterns is best for monitoring the return of diaphragmatic function?

      Your Answer: Post-tetanic count stimulation

      Explanation:

      Certain skeletal muscles are more resistant to the effects of neuromuscular blocking agents, both non-depolarizing and depolarizing. The diaphragm is the most resistant. The muscles of the larynx and the corrugator supercilii are less resistant. The abdominal, orbicularis oris, and limb peripheral muscles are the most sensitive muscles.

      Twitch stimulation patterns:

      Supramaximal single stimulus:

      The frequency ranges from 1 Hz to 0.1 Hz (one every second to one every 10 seconds)
      The response is proportional to the frequency of the event.
      It has limited clinical utility because it only tells you whether or not a patient is paralysed (no information on degree of paralysis).

      Over the course of 0.5 seconds (2 Hz), four supramaximal stimulate were applied:

      It is possible to see ‘fade’ and use it as a basis for evaluation.
      This stimulation pattern is used to determine the degree of blockade (1-2 twitches is appropriate for abdominal surgery)
      If the train of four (TOF) count is 1-2, reversal agents can be used in conjunction with medium-acting neuromuscular blocking agents.

      Ratio of TOF:

      This is the ratio of the 4th twitch amplitude to the 1st twitch amplitude.
      The ratio decreases with non-depolarising block and is inversely proportional to the degree of block, allowing objective measurement of residual neuromuscular blockade.
      To achieve adequate reversal, the ratio (as measured by accelerography) must be between 0.7 and 0.9.

      Count of twitches after a tetanic experience(PTC):

      50 Hz for 5 seconds, then a 3 second pause, followed by a single 1 Hz twitch stimulus.
      When the TOF count is zero, this stimulation pattern is used to assess deep blockade (that is, in neurosurgery, microsurgery or ophthalmic surgery when even small movements of a patient will disturb the surgical field)
      It gives an estimate of how long it will take for the response to return to single twitches, allowing assessment of blocks that are too deep for any other technique.
      A palpable post-tetanic count (PTC) of 2 indicates no twitch response for about 20-30 minutes, and a PTC of 5 indicates no twitch response for about 10-15 minutes.

      This is without a doubt the best way to keep track of paralysis in patients who need to avoid diaphragmatic movement. It’s best to use drug infusions and aim for a PTC of 2. After a tetanic stimulus, acetylcholine is mobilised, causing post-tetanic potentiation.

      Stimulation in Two Bursts:

      750 milliseconds between two short bursts of 50 Hz
      This stimulation pattern is used to assess small amounts of residual blockade manually (tactile).

    • This question is part of the following fields:

      • Clinical Measurement
      14.6
      Seconds
  • Question 7 - Arterial pressure waveforms give an indication of the operation of the heart and...

    Correct

    • Arterial pressure waveforms give an indication of the operation of the heart and the patient's clinical state.

      Which of the following listed characteristics of arterial waveforms is most indicative of myocardial contractility?

      Your Answer: Slope of the upstroke of the curve.

      Explanation:

      Arterial pressure waveforms is an invasive form of monitoring cardiac parameters. It provides a lot of information on the performance of the heart from different sections, including:

      Cardiac measurements:

      Heart rate
      Systolic pressure
      Diastolic pressure
      Mean arterial pressure
      Pulse pressure
      Change in pulse amplitude corresponding to respiratory changes
      Slope of anacrotic limb associated with aortic stenosis

      From the shape of the arterial waveform displayed:

      Slope of anacrotic limb represents aortic valve and LVOT flow
      Indications of aortic stenosis (AS): Slurred wave, collapsing wave
      Rapid systolic decline in LVOTO
      Bisferiens wave in HOCM
      Low dicrotic notch in states with poor peripheral resistance
      Position and quality of dicrotic notch as a reflection of the damping coefficient

      For this question, the upstroke slope of the pressure wave is indicative of myocardial contractility and is mathematically represented as:

      dP/dt, which represents a change of pressure with regards to time.

    • This question is part of the following fields:

      • Clinical Measurement
      31.8
      Seconds
  • Question 8 - The following are results of some pulmonary function tests:

    Measurement - Predicted result -...

    Incorrect

    • The following are results of some pulmonary function tests:

      Measurement - Predicted result - Test result
      Forced vital capacity (FVC) (btps) - 3.21 - 1.94
      Forced expiratory volume in 1 second (FEV1) (btps) - 2.77 - 1.82
      FEV1/FVC ratio % (btps) - 81.9 - 93.5
      Peak expiratory flow (PEF) (L/second) - 6.55 - 3.62
      Maximum voluntary ventilation (MVV) (L/minute) - 103 - 87.1

      Which statement applies to the results?

      Your Answer: These pulmonary function tests are normal

      Correct Answer: The patient has a moderate restrictive pulmonary defect

      Explanation:

      Severity of a reduction in restrictive defect (%FVC) or obstructive defect (%FEV1/FVC) predicted are classified as follows:

      Mild 70-80%
      Moderate 60-69%
      Moderately severe 50-59%
      Severe 35-49%
      Very severe <35% This patient has a %FVC predicted of 60.4% and this corresponds to a moderate restrictive deficit. %FEV1/FVC ratio is 93.5%. FEV1/FVC ratio 80% < predicted and VC < 80% = mixed picture. FEV1/FVC ratio 80% < predicted and VC > 80% = obstructive picture.

      FEV1/FVC ratio 80% > predicted and VC > 80% = normal picture.

      FEV1/FVC ratio 80% > predicted and VC < 80% predicted= restrictive picture. The integrity of the alveolar-capillary barrier is measured by carbon monoxide transfer factor (TLCO) and carbon monoxide transfer coefficient (KCO). These values are seen to be reduced in emphysema, interstitial lung diseases and in pulmonary vascular pathology. However, the KCO (as % predicted) is high in extrapulmonary restriction (pleural, chest wall and respiratory neuromuscular disease), and in loss of lung units provided the structure of the lung remaining is normal. The KCO distinguishes extrapulmonary (high KCO) causes of ‘restriction’ from intrapulmonary causes (low KCO).

    • This question is part of the following fields:

      • Clinical Measurement
      39
      Seconds
  • Question 9 - Which of the following lung parameters can be measured directly using spirometry? ...

    Incorrect

    • Which of the following lung parameters can be measured directly using spirometry?

      Your Answer: Residual volume

      Correct Answer: Vital capacity

      Explanation:

      Spirometry measures the total volume of air that can be forced out in one maximum breath, that is the total lung capacity (TLC), to maximal expiration, that is the residual volume (RV).

      It is conducted using a spirometer which is capable of measuring lung volumes using techniques of dilution.

      During spirometry, the following measurements can be determined:
      Forced vital capacity (FVC)/vital capacity (VC): The maximum volume of air exhaled in one single forced breathe.
      Forced expiratory volume in one second (FEV1)
      FEV1/FVC ratio
      Peak expiratory flow (PEF): the maximum amount of air flow exhaled in one blow.
      Forced expiratory flow (mid expiratory flow): the flow at 25%, 50% and 75% of FVC
      Inspiratory vital capacity (IVC): The maximum volume of air inhaled after a full total expiration.

      Anatomical dead space is measured using a single breath nitrogen washout called the Fowler’s method.

      Residual volume and total lung capacity are both measured using the body plethysmograph or helium dilution

      The functional residual capacity is usually measured using a nitrogen washout or the helium dilution technique.

    • This question is part of the following fields:

      • Clinical Measurement
      23.3
      Seconds
  • Question 10 - Which of the following options will cause an artificial increase in pulse oximeter...

    Correct

    • Which of the following options will cause an artificial increase in pulse oximeter (SpO2) readings?

      Your Answer: Heavy smoker

      Explanation:

      A pulse oximeter is a piece of medical equipment used as a non-invasive method of measuring the oxygen saturation of blood.

      It works by measuring the ratio of absorption of red and infrared light in a section of blood flow, as red light is largely absorbed by deoxygenated blood, and infrared light is largely absorbed by oxygenated blood.

      Pulse oximetry relies on photoplethysmography (PPG) waveforms. The oximeter has 2 sides, with different functions. One side houses light-emitting diodes which are responsible for transmitting 2 light wavelengths, 660nm for red light and 940nm for near infrared light. The other side is a photodetector. The light emitted travels through the body and the amount that is not absorbed is measured by the photodetector.

      Smokers often have increased levels of carboxy haemoglobin (COHb). This leads to artificial increases in pulse oximeter readings as it is unable to differentiate between COHb and oxyhaemoglobin (O2HB) as they both absorb red light at 660nm. Every 1% increase of circulating carboxyhaemoglobin, results in a correlative 1% increase in oximeter readings.

      Prilocaine toxicity will cause an artificial decrease in oximeter readings. This is because prilocaine metabolites cause methemoglobinemia (MetHB), which are dysfunctional haemoglobins unable to properly transport oxygen. In this case, a laboratory multiwavelength co-oximeter is recommended for a more accurate reading.

      Anaemia will not affect oximeter readings as long as haemoglobins in the blood are normal.

      Sickle cell disease does not affect oximeter readings despite its ability to cause hypoxia and shift the oxygen dissociation curve to the right.

      Brown-red fingernail polish will cause an underestimation of pulse oximeter readings.

    • This question is part of the following fields:

      • Clinical Measurement
      11.1
      Seconds
  • Question 11 - Which of the following can be measured directly using spirometry? ...

    Correct

    • Which of the following can be measured directly using spirometry?

      Your Answer: Vital capacity

      Explanation:

      Spirometry measures the total volume of air that can be forced out in one maximum breath, that is the total lung capacity (TLC), to maximal expiration, that is the residual volume (RV).

      It is conducted using a spirometer which is capable of measuring lung volumes using techniques of dilution.

      During spirometry, the following measurements can be determined:
      Forced vital capacity (FVC)/vital capacity (VC): The maximum volume of air exhaled in one single forced breathe.
      Forced expiratory volume in one second (FEV1)
      FEV1/FVC ratio
      Peak expiratory flow (PEF): the maximum amount of air flow exhaled in one blow.
      Forced expiratory flow (mid expiratory flow): the flow at 25%, 50% and 75% of FVC
      Inspiratory vital capacity (IVC): The maximum volume of air inhaled after a full total expiration.

      Anatomical dead space is measured using a single breath nitrogen washout called the Fowler’s method.

      Residual volume and total lung capacity are both measured using the body plethysmograph or helium dilution

      The functional residual capacity is usually measured using a nitrogen washout or the helium dilution technique.

    • This question is part of the following fields:

      • Clinical Measurement
      24.7
      Seconds
  • Question 12 - If the speed of an ultrasound in soft tissue is 1540 meters per...

    Correct

    • If the speed of an ultrasound in soft tissue is 1540 meters per second, what is the estimated wavelength produced if the frequency of a generated ultrasound is 10 megahertz?

      Your Answer: 0.15 millimetre

      Explanation:

      Wavelength can be computed as follows:

      Wavelength = velocity/frequency

      In the given problem, the values stated are:

      Frequency = 10 x 10^6
      Velocity = 1540 meters per second

      Wavelength = 1540/(10×10^6)
      Wavelength = 1540/10,000,000 meters
      Wavelength = 0.15 millimetres

    • This question is part of the following fields:

      • Clinical Measurement
      17.8
      Seconds
  • Question 13 - The statement that best describes temperature management is: ...

    Incorrect

    • The statement that best describes temperature management is:

      Your Answer: Thermistors use the Seebeck effect to measure temperature

      Correct Answer: Gauge thermometers use coils of different metals with different co-efficients of expansion which either tighten or relax with changes in temperature

      Explanation:

      There are different types of temperature measurement. These include:

      Thermistor – this is a type of semiconductor, meaning they have greater resistance than conducting materials, but lower resistance than insulating materials. There are small beads of semiconductor material (e.g. metal oxide) which are incorporated into a Wheatstone bridge circuit. As the temperature increases, the resistance of the bead decreases exponentially

      Thermocouple – Two different metals make up a thermocouple. Generally, in the form of two wires twisted, welded, or crimped together. Temperature is sensed by measuring the voltage. A potential difference is created that is proportional to the temperature at the junction (Seebeck effect)

      Platinum resistance thermometers (PTR) – uses platinum for determining the temperature. The principle used is that the resistance of platinum changes with the change of temperature. The thermometer measures the temperature over the range of 200°C to1200°C. Resistance in metals show a linear increase with temperature

      Tympanic thermometers – uses infrared radiation which is emitted by all living beings. It analyses the intensity and wavelength and then transduces the heat energy into a measurable electrical output

      Gauge/dial thermometers – Uses coils of different metals with different co-efficient of expansion. These either tighten or relax with changes in temperature, moving a lever on a calibrated dial.

    • This question is part of the following fields:

      • Clinical Measurement
      1017.9
      Seconds
  • Question 14 - An arterial pressure transducer is supposedly in direct correlation to change, thus it...

    Correct

    • An arterial pressure transducer is supposedly in direct correlation to change, thus it is dependent on zero gradient drift and zero offset. Which of the following values will best compensate for the gradient drift?

      Your Answer: 0 mmHg and 200 mmHg

      Explanation:

      Since an arterial pressure transducer, and every other measuring apparatus, is prone to errors due to offset and gradient drifts, regular calibration is required to maintain accuracy of the instrument. The two-point calibration pressure values of 0 mmHg and 200 mmHg are within the physiologic range and can best compensate for the gradient drift.

    • This question is part of the following fields:

      • Clinical Measurement
      6.3
      Seconds
  • Question 15 - Which of these statements is true about spirometry? ...

    Correct

    • Which of these statements is true about spirometry?

      Your Answer: A capacity is the sum of two or more volumes

      Explanation:

      Functional residual capacity (FRC) is 1.7 to 3.5L/kg

      A capacity is the sum of two or more volumes. The total lung capacity (TLC) is total sum of the volume of gas present in all lung compartments upon maximum inspiration. It is represented mathematically as:

      Total lung capacity (TLC) = Vital capacity (VC) + Residual volume (RV)

      The residual volume (RV) is the volume of gas still present within the lung post maximum exhalation. It cannot be measured by spirometry, but can be using a body plethysmograph and also with the helium dilution technique.

      Closing capacity (CC) is the volume of gas within the lungs at which small airways close upon expiration. It increases with age and is especially important when it surpasses the FRC as it causes changes in ventilation/perfusion mismatch and hypoxia.
      In the supine position, a patient with a normal body mass index and no history of lung pathology, the CC equals the FRC at approximately 44, and at approximately 66 at standing position.

    • This question is part of the following fields:

      • Clinical Measurement
      30.8
      Seconds
  • Question 16 - A 23-year-old man, has just undergone surgery under general anaesthesia. He has experienced...

    Correct

    • A 23-year-old man, has just undergone surgery under general anaesthesia. He has experienced a severe reaction to the anaesthetic agent resulting in malignant hyperthermia (MH) for which he has been referred for treatment.

      What investigation can be conducted to determine a patient's susceptibility to malignant hyperthermia?

      Your Answer: In vitro muscle contraction test using caffeine

      Explanation:

      Malignant hyperthermia (MH) is a autosomal dominant inherited medical condition which predisposes affected individuals to a clinical syndrome of hypermetabolism which involves abnormal ryanodine receptors in skeletal muscle causing a deregulation of calcium in muscle.

      It is a life threatening condition requiring immediate medical intervention. It often lies dormant until triggered in susceptible individuals mostly by volatile inhaled anaesthetic agents and succinylcholine which is a muscle relaxant.

      The signs and symptoms of MH are related to this hypermetabolism, which includes an increase in carbon dioxide production, metabolic and respiratory acidosis, accelerated oxygen consumption, heat production, activation of the sympathetic nervous system, hyperkalaemia, disseminated intravascular coagulation (DIC), and multiple organ dysfunction and failure.

      Early signs of MH to look out for in patients includes an uptick in end-tidal carbon dioxide (even with increasing minute ventilation), tachycardia, muscle rigidity, tachypnoea, and hyperkalaemia. Later signs include fever, myoglobinuria, and multiple organ failure.

      In vitro muscle contracture test (IVCT) is the standard for determining individual susceptibility to MH. It is conducted by measuring the force of muscle contraction after exposing the patient’s muscle sample to halothane and caffeine., the sample is normally taken from the vastus medialis or lateralis under regional anaesthesia.

    • This question is part of the following fields:

      • Clinical Measurement
      24.1
      Seconds
  • Question 17 - A 27-year-old woman arrives at the emergency room after intentionally ingesting 2 g...

    Correct

    • A 27-year-old woman arrives at the emergency room after intentionally ingesting 2 g of amitriptyline.

      A Glasgow coma score of 6 was discovered, as well as a pulse rate of 140 beats per minute and a blood pressure of 80/50 mmHg.

      Which of the following ECG changes is most likely to indicate the onset of life-threatening arrhythmias?

      Your Answer: Prolongation of the QRS complex

      Explanation:

      Arrhythmias and/or hypotension are the most common causes of death from tricyclic antidepressant (TCA) overdose.

      The quinidine-like actions of tricyclic antidepressants on cardiac tissues are primarily responsible for their toxicity. Conduction through the His-Purkinje system and the myocardium slows as phase 0 depolarisation of the action potential slows. QRS prolongation and atrioventricular block are caused by slowed impulse conduction, which also contributes to ventricular arrhythmias and hypotension.

      Arrhythmias can also be caused by abnormal repolarization, impaired automaticity, cholinergic blockade, and inhibition of neuronal catecholamine uptake, among other things.

      Acidaemia, hypotension, and hyperthermia can all exacerbate toxicity.

      The anticholinergic effects of tricyclic antidepressants, as well as the blockade of neuronal catecholamine reuptake, cause sinus tachycardia. Sinus tachycardia is usually well tolerated and does not require treatment. It can be difficult to tell the difference between sinus tachycardia and ventricular tachycardia with QRS prolongation.

      A QRS duration of more than 100 milliseconds indicates a higher risk of arrhythmia and should be treated with systemic sodium bicarbonate.

      The tricyclic is dissociated from myocardial sodium channels by serum alkalinization, and the extracellular sodium load improves sodium channel function.

    • This question is part of the following fields:

      • Clinical Measurement
      130.2
      Seconds
  • Question 18 - A patient has a myocardial infarction with anterior ST elevation. There are the...

    Correct

    • A patient has a myocardial infarction with anterior ST elevation. There are the following observations:

      Cardiac output 2.0 L/min
      Blood pressure 80/60 mmHg
      CVP 20 mmHg
      SpO2 91% on 4 L/min oxygen

      What is the most logical physiological explanation for these findings?

      Your Answer: Biventricular failure

      Explanation:

      The occlusion of the left anterior descending (LAD) coronary artery causes anterior ST elevation myocardial infarction (STEMI). It has the worst prognosis of all the infarct locations due to its larger infarct size. It has a higher rate of total mortality (27 percent versus 11 percent), heart failure (41 percent versus 15 percent), and a lower ejection fraction on admission than an inferior myocardial infarction (38 percent versus 55 percent ).

      The LAD artery supplies the majority of the interventricular septum, as well as the anterior, lateral, and apical walls of the left ventricle, as well as the majority of the right and left bundle branches and the bicuspid valve’s anterior papillary muscle (left ventricle).

      The left or right ventricle’s end-diastolic volume (EDV) is the volume of blood in each chamber at the end of diastole before systole. Preload is synonymous with the EDV.

      120 mL is a typical left ventricular EDV (range 65-240 mL). The EDV of the right ventricle in a typical range is (100-160 mL).

      With an ejection fraction (EF) of less than 45 percent, the patient is most likely suffering from systolic dysfunction. Increases in right and left ventricular end-diastolic pressures and volumes are likely with a reduced EF because the ventricles are not adequately emptied. The left atrium and the pulmonary vasculature are affected by the increased pressures on the left side of the heart.

      By causing an imbalance of the Starling forces acting across the capillaries, increased hydrostatic pressure in the pulmonary circulation favours the development of pulmonary oedema. With cardiogenic pulmonary oedema, capillary permeability is likely to remain unchanged.

      Biventricular failure will result as a result of the pressure changes being transmitted to the right side of the circulation. The patient’s systemic vascular resistance is likely to be elevated as well, but it is not the most likely cause of his symptoms. The patient is suffering from cardiogenic shock as a result of biventricular failure. The patient has low cardiac output and is hypotensive. Right ventricular filling pressures are elevated, indicating right ventricular dysfunction.

    • This question is part of the following fields:

      • Clinical Measurement
      15.3
      Seconds
  • Question 19 - A 50-year old man for septoplasty has a BMI of 32kg/m2 and neck...

    Correct

    • A 50-year old man for septoplasty has a BMI of 32kg/m2 and neck circumference of 44 cm. He is troubled by intermittent nasal obstruction which causes his partner sleeps in a separate bedroom because of snoring. He currently on ramipril for hypertension.

      Which of the following best approximates to his STOP-BANG score from the information above?

      Your Answer: 5

      Explanation:

      The STOP-BANG questionnaire is used to screen patients for obstructive sleep apnoea (OSA).

      The scoring system assigns one point for each feature.

      S: Snoring (louder than talking or loud enough to be heard through closed doors)
      T: Feeling tired, fatigued, or sleepy during daytime
      O: Observed apnoeas during sleep
      P: Hypertension
      B: BMI more than 35 kg/m2
      A: Age 50-years of age or greater
      N: Neck circumference (male 17 inches / 43cm or greater and female 16 inches / 41 or greater)
      G: Gender: Male

      Our patient has a score of 5 ( O, P, A, N, G)

      The score helps clinicians stratify patients for unrecognized OSA and target appropriate clinical management. It can also help triage patients for further investigation. A STOP-BANG score of 5-8 will identify patients with high probability of moderate to severe OSA in the surgical population.

    • This question is part of the following fields:

      • Clinical Measurement
      6.4
      Seconds
  • Question 20 - Which of the following statements about closing capacity is true? ...

    Incorrect

    • Which of the following statements about closing capacity is true?

      Your Answer: Is measured using Fowler's method

      Correct Answer: It is less than the functional residual capacity in a 30-year-old

      Explanation:

      Closing capacity refers to volume of gas within the lungs at which the conducting small airways begin to close, that is, the point during expiration when small airways close.

      It is calculated mathematically as:

      Closing capacity = Closing volume (CV) + Residual volume (RV)

      Functional residual capacity (FRC) is the volume of gas still present within the lungs post expiration.

      Closing capacity is lower than the functional residual capacity in younger adults, but begins to rise to eventually equal, and then exceed it with increasing age (at about middle age), increasing intrabdominal pressure, decreasing blood flow in the pulmonary system and parenchymal disease within the pulmonary system.

    • This question is part of the following fields:

      • Clinical Measurement
      13.9
      Seconds
  • Question 21 - The statement that best describes the classification of theatre equipment in terms of...

    Correct

    • The statement that best describes the classification of theatre equipment in terms of electrical safety is:

      Your Answer: A floating circuit is equipment applied to patient that is isolated from all its other parts

      Explanation:

      There are different classes of electrical equipment that can be classified in the table below:

      Class 1 – provides basic protection only. It must be connected to earth and insulated from the mains supply

      Class II – provides double insulation for all equipment. It does not require an earth.

      Class III – uses safety extra low voltage (SELV) which does not exceed 24 V AC. There is no risk of gross electrocution but risk of microshock exists.

      Type B – All of above with low leakage currents (0.5mA for Class IB, 0.1 mA for Class IIB)

      Type BF – Same as with other equipment but has ‘floating circuit’ which means that the equipment applied to patient is isolated from all its other parts.

      Type CF – Class I or II equipment with ‘floating circuits’ that is considered to be safe for direct connection with the heart. There are extremely low leakage currents (0.05mA for Class I CF and 0.01mA for Class II CF)

    • This question is part of the following fields:

      • Clinical Measurement
      12.7
      Seconds
  • Question 22 - Which statement is the most accurate when describing electrical equipment and shock? ...

    Correct

    • Which statement is the most accurate when describing electrical equipment and shock?

      Your Answer: Type CF is considered to safe for direct connection with the heart

      Explanation:

      There are different classes of electrical equipment that can be classified in the table below:

      Class 1 – provides basic protection only. It must be connected to earth and insulated from the mains supply

      Class II – provides double insulation for all equipment. It does not require an earth.

      Class III – uses safety extra low voltage (SELV) which does not exceed 24 V AC. There is no risk of gross electrocution but risk of microshock exists.

      Type B – All of above with low leakage currents (0.5mA for Class IB, 0.1 mA for Class IIB)

      Type BF – Same as with other equipment but has ‘floating circuit’ which means that the equipment applied to patient is isolated from all its other parts.

      Type CF – Class I or II equipment with ‘floating circuits’ that is considered to be safe for direct connection with the heart. There are extremely low leakage currents (0.05mA for Class I CF and 0.01mA for Class II CF)

    • This question is part of the following fields:

      • Clinical Measurement
      15.7
      Seconds
  • Question 23 - Which one of the following patients presenting for elective surgery has an American...

    Correct

    • Which one of the following patients presenting for elective surgery has an American Society of Anaesthesiologists (ASA) preoperative physical status grading of III?

      Your Answer: A 50-year old man with a BMI of 41 with a reduced exercise tolerance

      Explanation:

      The ASA physical status classification system is a system for assessing the fitness of patients before surgery. It was last updated in October 2014.

      ASA I A normal healthy patient
      ASA II A patient with mild systemic disease
      ASA III A patient with severe systemic disease
      ASA IV A patient with severe systemic disease that is a constant threat to life
      ASA V A moribund patient who is not expected to survive without the operation
      ASA VI A declared brain-dead patient whose organs are being removed for donor purposes

      A 20-year old woman who is 39-weeks pregnant with no other medical conditions – ASA II

      A 35-year-old man with a BMI of 29 with a good exercise tolerance who smokes-ASA II

      A 50-year old man with a BMI of 41 with a reduced exercise tolerance -ASA III

      A 65-year old woman with a BMI of 34 with treated hypertension with no functional limitations-ASA II

      A 73-year old man who has had a TIA ten-weeks ago but has a good exercise tolerance and is a non-smoker-ASA IV

    • This question is part of the following fields:

      • Clinical Measurement
      22.1
      Seconds
  • Question 24 - The following haemodynamic data is available from a patient with pulmonary artery catheter...

    Correct

    • The following haemodynamic data is available from a patient with pulmonary artery catheter inserted:

      Pulse rate - 100 beats per minute
      Blood pressure - 120/70mmHg
      Mean central venous pressure (MCVP) - 10mmHg
      Right ventricular pressure (RVP) - 30/4 mmHg
      Mean pulmonary artery wedge pressure (MPAWP) - 12mmHg

      Which value best approximates the patient's coronary perfusion pressure?

      Your Answer: 58mmHg

      Explanation:

      Coronary perfusion pressure(CPP), the difference between aortic diastolic pressure (Pdiastolic) and the left ventricular end-diastolic pressure (LVEDP), is mainly determined by the formula:

      CPP = Pdiastolic -LVEDP
      where
      Pdiastolic is the lowest pressure in the aorta before left ventricular ejection and
      LVEDP is measured directly during a cardiac catheterisation or indirectly using a pulmonary artery catheter. The pulmonary artery occlusion or wedge pressure approximates best with LVEDP.

      Using this patient’s haemodynamic data:

      CPP = Pdiastolic – MPAWP
      COO = 70 – 12 = 58mmHg

    • This question is part of the following fields:

      • Clinical Measurement
      18.7
      Seconds
  • Question 25 - Which of the following is true about the bispectral index (BIS)? ...

    Incorrect

    • Which of the following is true about the bispectral index (BIS)?

      Your Answer: Burst suppression patterns are evident at levels below 65

      Correct Answer: Sevoflurane lowers BIS more than ketamine

      Explanation:

      The bispectral index (BIS) monitors works to determine the level of consciousness of a patient by processing electroencephalographic (EEG) signals to obtain a value between 0 and 100, where 0 reflects no brain activity, and 100 reflects a patient is completely awake.

      The general meaning of BIS values are:

      >95: Patient is in an awake state.
      65-85: Patient is in a sedated state.
      40-65: Patient is in a state that is optimal for general surgery.
      <40: Patient is in a deep hypnotic state It is important in measuring the depths of anaesthesia to prevent haemodynamic changes or patient awareness during surgery. The nature of anaesthetic agent used is a determinant factor in resultant BIS values. Intravenous agents, such as propofol, thiopental and midazolam, result in a deeper hypnotic state, whilst inhalation agents have a lesser hypnotic effect at the same BIS values. Certain agents result in inaccurate BIS values such as ketamine and nitrous oxide (NO). These two agents appear to increase the BIS value, whilst putting the patient in a deeper hypnotic state, and should therefore not be used with BIS monitoring. Hypothermia also affects the BIS value as it causes a 1.12 per °C decrease in body temperature.

    • This question is part of the following fields:

      • Clinical Measurement
      101.6
      Seconds
  • Question 26 - A 28-year-old man is admitted to the critical care unit. He has been...

    Incorrect

    • A 28-year-old man is admitted to the critical care unit. He has been diagnosed with adult respiratory distress syndrome and is being ventilated. His haemodynamic condition is improved using a pulmonary artery flotation.

      His readings are listed below:

      Haemoglobin concentration: 10 g/dL
      Mixed venous oxygen saturation: 70%
      Mixed venous oxygen tensions (PvO2): 50 mmHg

      Estimate his mixed venous oxygen content (mL/100mL).

      Your Answer: 10.5

      Correct Answer: 9.5

      Explanation:

      Mixed venous oxygen content (CvO2) is the oxygen concentration in 100mL of mixed venous blood taken from the pulmonary artery. It is usually 12-17 mL/dL (70-75%). It is represented mathematically as:

      CvO2 = (1.34 x Hgb x SvO2 x 0.01) + (0.003 x PvO2)

      Where,

      1.34 = Huffner’s constant
      Hgb = Haemoglobin level (g/dL)
      SvO2 = % oxyhaemoglobin saturation of mixed venous blood
      PvO2 = 0.0225 = mL of O2 dissolved per 100mL plasma per kPa, or 0.003 mL per mmHg

      Therefore,

      CvO2 = (1.34 x 10 x 70 x 0.01) + (0.003 x 50)

      CvO2 = 9.38 + 0.15 = 9.53 mL/100mL

    • This question is part of the following fields:

      • Clinical Measurement
      212.8
      Seconds
  • Question 27 - A 49-year-old woman is admitted to hospital. She is scheduled for surgery and...

    Correct

    • A 49-year-old woman is admitted to hospital. She is scheduled for surgery and is undergoing preoperative assessment.

      As part of the preoperative assessment, her functional capacity is estimated. At 50kg in weight, she is able to cycle along a flat surface at a speed of 10-14 miles/hour (8 metabolic equivalents or METs).

      Provide the best estimated value of oxygen consumption (VO2) for eight METs.

      Your Answer: 1400 mL/minute

      Explanation:

      Oxygen consumption (VO2) refers to the optimal amount of oxygen used by the body during exercise.

      It is calculated mathematically by:

      VO2 = 3.5 x 50 x 8 = 1400 mL/kg/minute

      where,

      1 MET = 3.5 mL O2/kg/minute is utilized by the body.

      Note:

      1 MET Eating
      Dressing
      Use toilet
      Walking slowly on level ground at 2-3 mph
      2 METs Playing a musical instrument
      Walking indoors around house
      Light housework
      4 METs Climbing a flight of stairs
      Walking up hill
      Running a short distance
      Heavy housework, scrubbing floors, moving heavy furniture
      Walking on level ground at 4 mph
      Recreational activity, e.g. golf, bowling, dancing, tennis
      6 METs Leisurely swimming
      Leisurely cycling along the flat (8-10 mph)
      8 METs Cycling along the flat (10-14 mph)
      Basketball game
      10 METs Moderate to hard swimming
      Competitive football
      Fast cycling (14-16 mph)

    • This question is part of the following fields:

      • Clinical Measurement
      21.3
      Seconds
  • Question 28 - A 50-year-old man is admitted in hospital. Over four hours, he produces 240...

    Correct

    • A 50-year-old man is admitted in hospital. Over four hours, he produces 240 mL of urine and has a plasma creatinine concentration is 10 mcg/mL. The normal concentration of creatinine in urine is 1.25 mg/mL.

      Calculate his approximate creatinine clearance.

      Your Answer: 125 ml/minute

      Explanation:

      Creatinine clearance is a test used to approximate the glomerular filtration rate (GFR) as an assessment of kidney function.

      Creatinine is formed during the breakdown of dietary sources of meat and skeletal muscle. It is secreted at a consistent concentration and pace into the body’s circulation, and is easily filtered across the glomerulus without being reabsorbed or metabolized by the kidney.

      It is represented mathematically as:
      Creatinine clearance (CL) = U x V/P
      where,
      U: Urinary creatinine concentration (mg/mL)
      V: Volume of urine (mL/min)
      P: Plasma creatinine concentration (mg/mL)

      Therefore, in this case:
      CL: 1.25 x 1 = 125mL/min
      0.1

    • This question is part of the following fields:

      • Clinical Measurement
      10.3
      Seconds
  • Question 29 - A 76-year-old female, presents to her GP. She complains of fatigue and increased...

    Correct

    • A 76-year-old female, presents to her GP. She complains of fatigue and increased shortness of breath. On examination, she is noted to have pallor, an increased respiratory and heart rate. Her GP requests further diagnostic investigations, including a full blood count (FBC) which finds decreased MCV and MCHC.

      What is the most likely cause of her symptoms?

      Your Answer: Iron deficiency

      Explanation:

      The patient’s diagnosis is microcytic hypochromic anaemia which is often as a result of iron deficiency and thalassaemia.

      Macrocytic anaemia is often caused by folate and B12 deficiencies and alcohol abuse.

      Normocytic normochromic anaemia is often caused by acute blood loss, haemolytic anaemia, anaemia of chronic disease and leucoerythroblastic anaemias.

    • This question is part of the following fields:

      • Clinical Measurement
      13.3
      Seconds
  • Question 30 - With regards to devices for temperature management, all of these are used EXCEPT:...

    Incorrect

    • With regards to devices for temperature management, all of these are used EXCEPT:

      Your Answer: Thermocouple works using the potential difference which is created when dissimilar metals are joined

      Correct Answer: Thermistors use the resistance of a semiconductor bead which increases exponentially as the temperature increases

      Explanation:

      There are different types of temperature measurement. These include:

      Thermistor – this is a type of semiconductor, meaning they have greater resistance than conducting materials, but lower resistance than insulating materials. There are small beads of semiconductor material (e.g. metal oxide) which are incorporated into a Wheatstone bridge circuit. As the temperature increases, the resistance of the bead decreases exponentially

      Thermocouple – Two different metals make up a thermocouple. Generally, in the form of two wires twisted, welded, or crimped together. Temperature is sensed by measuring the voltage. A potential difference is created that is proportional to the temperature at the junction (Seebeck effect)

      Platinum resistance thermometers (PTR) – uses platinum for determining the temperature. The principle used is that the resistance of platinum changes with the change of temperature. The thermometer measures the temperature over the range of 200°C to1200°C. Resistance in metals show a linear increase with temperature

      Tympanic thermometers – uses infrared radiation which is emitted by all living beings. It analyses the intensity and wavelength and then transduces the heat energy into a measurable electrical output

      Gauge/dial thermometers – Uses coils of different metals with different co-efficient of expansion. These either tighten or relax with changes in temperature, moving a lever on a calibrated dial.

    • This question is part of the following fields:

      • Clinical Measurement
      18.5
      Seconds

SESSION STATS - PERFORMANCE PER SPECIALTY

Clinical Measurement (21/30) 70%
Passmed