-
Question 1
Incorrect
-
Which one of the following statement is true regarding United Kingdom gas cylinders?
Your Answer: The nominal gauge pressure of full nitrous oxide cylinder at 20°Cis 13700 kPa
Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred
Explanation:Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.
At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.
Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.
The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Incorrect
-
From the following statements, which is true of humidification?
Your Answer: The wet and dry bulb hygrometer measures absolute humidity
Correct Answer: Ultrasonic humidifier can achieve greater than 100% relative humidity
Explanation:Increasing temperature increases the amount of water vapour contained in air; for example, at 20°C, air contains about 17 g/m3, and at 37°C, air contains about 44 g/m3. The wet and dry bulb hygrometer, like the hair hygrometer, measures relative humidity.
Under normal operating conditions, Heat and moisture exchangers (HMEs) allows relative humidity of up to 70% to be achieved. Mucus can impair their performance, and they should not be used for longer than 24 hours.
Hot water bath humidifiers might cause scalding, condensed water in the tubing can interfere with gas flow, and there is a danger of infection.
The ultrasonic humidifier operates at roughly 2 MHz and may attain relative humidity levels much above 100%.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Incorrect
-
Which of the following statements is true about oxygen face masks?
Your Answer: Variable performance devices are colour coded and are marked with the recommended oxygen flow rate
Correct Answer: The oxygen concentration delivered by high air flow oxygen enrichment devices is not dependent on the respiratory pattern of the patient
Explanation:The normal peak inspiratory flow in healthy individuals is 20-30 L/min during each normal tidal ventilation. This is expected to increase with greater respiratory rate and deeper inspiration.
Face masks are used to facilitate the delivery of oxygen from a breathing system to a patient. Face masks can be divided into two types: fixed performance or variable performance devices.
In fixed performance devices (also known as high air flow oxygen enrichment or HAFOE), fixed inspired oxygen concentration is delivered to the patent, independent and greater than that of the patient’s peak inspiratory flow rate (PIFR). No random entrainment is expected to occur at the time of PIFR, hence, the oxygen concentration in HAFOE devices is not dependent on the patient’s respiratory pattern.
Moreover, in HAFOE masks, the concentration of oxygen at a given oxygen flow rate is determined by the size of the constriction; a device with a greater entrainment aperture delivers a lower oxygen concentration. Therefore, a 40% Venturi device will have lesser entrainment aperture when compared to a 31% Venturi. Venturi masks allow relatively fixed concentrations of supplemental oxygen to be inspired e.g. 24%, 28%, 31%, 35%, 40% and 60% oxygen. These are colour coded and marked with the recommended oxygen flow rate.
Variable performance devices deliver variable inspired oxygen concentration to the patient, and is dependent on the PIFR. The PIFR can often exceed the flow rate at which oxygen or an oxygen/air mixture is supplied by the device, depending on a patient’s inspiratory effort. In addition, these masks allow expired air to be released through the holes in the sides of the mask. Thus, with increased respiratory rate, rebreathing of alveolar gas from inside the mask may occur.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Correct
-
The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients.
Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?Your Answer: II (CF)
Explanation:Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Correct
-
The half-empty cylinder weighs 4.44 kg.
The tare weight of a nitrous oxide cylinder is 4 kg.
The molecular weight of nitrous oxide is 44gm.
Based on the data, how many litres of nitrous oxide remains in the cylinder for use?Your Answer: 224 litres
Explanation:The Tare weight of a cylinder is the weight when it is empty. So,
Weight of cylinder – tare weight = weight of remaining N2O (g).
4.44 kg – 4 kg = 0.44 kg
Here,
0.44 kg of nitrous oxide remains in the cylinderSince the molecular weight of nitrous oxide is 44 g and one mole of an ideal gas will occupy a volume of 22.4 litres at STP
Therefore amount left in the cylinder is several (gN2O/44) x 22.4 litres of N2O.(440/44) x 22.4 = 224 litres.
-
This question is part of the following fields:
- Basic Physics
-
-
Question 6
Incorrect
-
An emergency appendicectomy is being performed on a 20 year old man. For maintenance of anaesthesia, he is being ventilated using a circle system with a fresh gas flow (FGF) of 1 L/min (air/oxygen and sevoflurane). The trace on the capnograph shows a normal shape.
The table below demonstrates the changes in the end-tidal and baseline carbon dioxide measurements of the capnograph at 10 and 20 minutes of anaesthesia maintenance.
End-tidal CO2: 4.9 kPa vs 8.4kPa (10 minutes vs 20 minutes)
Baseline end-tidal CO2: 0.2 kPa vs 2.4kPa
Pulse 100-107 beats per minute, systolic blood pressure 125-133 mmHg and oxygen saturation 98-99%.
Which of the following is the single most important immediate course of action?Your Answer: Replace the soda lime
Correct Answer: Increase the FGF
Explanation:End-tidal carbon dioxide (ETCO2) monitoring has been an important factor in reducing anaesthesia-related mortality and morbidity. Hypercarbia, or hypercapnia, occurs when levels of CO2 in the blood become abnormally high (Paco2 >45 mm Hg). Hypercarbia is confirmed by arterial blood gas analysis. When using capnography to approximate Paco2, remember that the normal arterial–end-tidal carbon dioxide gradient is roughly 5 mm Hg. Hypercarbia, therefore, occurs when PETco2 is greater than 40 mm Hg.
The most likely explanation for the changes in capnograph is either exhaustion of the soda lime and a progressive rise in circuit dead space.
Inspect the soda lime canister for a change in colour of the granules. To overcome soda lime exhaustion, the first step is to increase the fresh gas flow (FGF) (Option A). Then, if need arises, replace the soda lime granules. Other strategies that can work are changing to another circuit or bypassing the soda lime canister, but remember that both these strategies are employed only after increasing FGF first. Exclude other causes of equipment deadspace too.
There are also other causes for hypercarbia to develop intraoperatively:
1. Hypoventilation is the most common cause of hypercapnia. A. Inadequate ventilation can occur with spontaneous breathing due to drugs like anaesthetic agents, opioids, residual NMDs, chronic respiratory or neuromuscular disease, cerebrovascular accident.
B. In controlled ventilation, hypercapnia due to circuit leaks, disconnection or miscalculation of patient’s minute volume.
2. Rebreathing – Soda lime exhaustion with circle, inadequate fresh gas flow into Mapleson circuits and increased breathing system deadspace.
3. Endogenous source – Tourniquet release, hypermetabolic states (MH or thyroid storm) and release of vascular clamps.
4. Exogenous source – Absorption of CO2 from pneumoperitoneum. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Incorrect
-
Regarding laminar gas flow, which of the following options has the most influence on laminar flow?
Your Answer: Coefficient of viscosity
Correct Answer: Diameter of tube
Explanation:Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:
1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
2. Few directional changes
3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
5. Double flow by doubling pressure as long as the flow pattern remains laminarPoiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Correct
-
Fixed performance devices like high air flow oxygen enrichment (HAFOE) masks have large volumes of air entrained into a flow of 100% oxygen.
The term that best describes the physics behind air entrainment is?Your Answer: Bernoulli's principle
Explanation:Bernoulli’s principle states that as the speed of a moving fluid increases, there is a simultaneously decrease in static pressure or a decrease in the fluid’s potential energy.
This is seen in the simultaneous increase in speed and kinetic energy and fall in pressure that causes entrainment of large volumes of air into a flow of 100% oxygen in the nozzle of HAFOE masks.The reduction in fluid pressure that happens when a fluid flows through a constriction in a tube is the Venturi effect.
When a flow of gas or liquid attaches itself to a nearby surface and remains attached even when the surface curves away from the initial direction of flow, this is the Coanda effect.
The branch of engineering and technology that is concerned with the building of devices that use the flow and pressure of a fluid for functions usually performed by electronic devices is Fluidics . Fluidic logic is used to power some ventilators.
The branch of engineering that utilises pressurised gases is Pneumatics.
-
This question is part of the following fields:
- Basic Physics
-
-
Question 9
Correct
-
A 74-year-old with a VVI pacemaker is undergoing a hip replacement.
Which of the following is most likely to predispose him to an electrical hazard?Your Answer: Use of cutting unipolar diathermy
Explanation:A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).
EMI has the potential to cause the following: Inhibition of pacing
Asynchronous pacing
Reset to backup mode
Myocardial burns, and
Trigger VF.Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.
Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.
Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.
Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).
A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Correct
-
General anaesthesia is administered to a patient in a hospital in Lhasa which is one of the highest cities in the world (at 11,975 feet). An Anaesthetic rotameter is normally calibrated at 20 C and 1 bar pressure and is known to be underread at altitude. The temperature of the theatre was 10 C.
Which one of the following physical properties is responsible for the rotameter inaccuracy in these conditions?Your Answer: Density of the gas
Explanation:Since the gas is less dense at higher altitudes, the density of a gas influences flows when passing through the orifice. Due to this reason, for a given flow rate, the bobbin will not be forced as far up the rotameter tube.
At higher altitudes, the volume of a fixed mass of gas increases, and therefore the molecules of gas are widely spaced resulting in a decrease in density with an increase in altitude.
Viscosity is simply termed as friction of gas. The viscosity of a gas is important only at low flow rates when the flow characteristic of the gas is laminar.
Charle’s law stated that the volume occupied by a fixed amount of gas is directly proportional to its absolute temperature (T) provided the pressure remains constant.
Boyle’s law for a fixed amount of gas at constant temperature, the pressure (P) and volume (V) are inversely proportional.
-
This question is part of the following fields:
- Basic Physics
-
-
Question 11
Correct
-
Which of the following derived SI units is correctly expressed as their base units?
Your Answer: Volt: m2.kg.s-3.A-1
Explanation:The following units are derived SI units of measurement.
Energy or work: kg.m2.s-2
The Joule (J) is the energy transferred to an object when a force of one newton acts on that object in the direction of its motion through a distance of one meter or N.m.Power: kg.m2.s-3
The Watt (W) = rate of transfer of energy or Joule per second J/s.Force: kg.m.s-2
One Newton (N) which is the international unit of measure for force = 1 kilogram meter per second squared. 1 Newton of force is the force required to accelerate an object with a mass of 1 kilogram 1 meter per second per second.Volt: kg.m2.s-3.A-1
The volt (V) is defined as the potential difference across a conductor when a current of one ampere dissipates one watt of power or W/A.Pressure: kg.m-1.s-2
A pascal (Pa) is force per unit area or N/m2. -
This question is part of the following fields:
- Basic Physics
-
-
Question 12
Correct
-
Concerning the physical principles of temperature measurement by a thermocouple, which of the following best describes it?
Your Answer: The bimetallic strip has a junction potential proportional to temperature
Explanation:A thermocouple, or a thermal junction, is temperature measuring device consisting of a pair of dissimilar metal (bimetallic) wires or strips joined together. Typically, copper and constantan (an alloy of 55% copper and 45% nickel) are used. When there is contact between these metals, a small voltage is generated in the order of millivolts. The magnitude of the thermojunction electromotive force (emf) is proportional to applied temperature (the Seebeck effect). This physical principle is applied in the measurement of temperature. The electromotive force at the measuring junction is proportional to temperature.
Two wires with different coefficients of expansion, joined together, can be used as a switch for thermostatic control.
Semiconductors are NOT used in thermocouple. The resistance of the measuring junction of a thermocouple is irrelevant.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 13
Correct
-
Which of the following is the smallest value of pressure?
Your Answer: 14.69 psi
Explanation:The SI unit of pressure is the pascal (Pa) and it is equal to one newton (N) per square meter (m2) or N/m2.
1 atmosphere (atm) is the equivalent of:
101325 Pa760 mmHg
1.01325 bar
1033.23 cmH2O.
14.69 pounds per square inch (psi)
1013.25 millibar (mbar) or hectopascals (hPa), and14.69 psi is equal to one atmosphere. The other values are equal to two atmospheres of pressure.
-
This question is part of the following fields:
- Basic Physics
-
-
Question 14
Incorrect
-
Which of the following statements is TRUE regarding an epidural set?
Your Answer: The filter has a 0.55 micron mesh
Correct Answer: 19G Tuohy needles have 0.5 cm markings
Explanation:A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings
18G Tuohy catheters are generally 9 to 10cm to hub
Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura
Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure
Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings)
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 15
Incorrect
-
When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?
Your Answer: Has a frequency of 500 KHz - 1 MHz
Correct Answer: Has a power output of up to 140 joules per second
Explanation:Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).
The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.
A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.
High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).
The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 16
Incorrect
-
Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?
Your Answer: Utilises a double cuff oscillotonometric system
Correct Answer: The cuff should be positioned at the same level as the heart
Explanation:Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.
The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.
The manual BP device is still the gold standard for BP measurement and monitoring.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 17
Correct
-
Very large SI units are easily expressed using mathematical prefixes.
One terabyte is equal to which of the following numbers?Your Answer: 1,000,000,000,000 bytes
Explanation:To denote large measured units, the following SI mathematical prefixes are used:
1 deca = 10 bytes (101)
1 hecto (h) = 100 bytes
1 kilo (k)= 1,000 bytes
1 mega (M) = 1,000,000 bytes
1 giga (G) = 1,000,000,000 bytes
1 Tera (T) = 1,000,000,000,000 bytes
1 Peta (P) = 1,000,000,000,000,000 bytes -
This question is part of the following fields:
- Basic Physics
-
-
Question 18
Incorrect
-
Regarding a paramagnetic oxygen analyser, the following statements are TRUE:
Your Answer: Can measure gases dissolved in blood
Correct Answer: Utilises null deflection
Explanation:The electrons in the outer shell of an oxygen molecule are unpaired, thus it has paramagnetic properties and is attracted into a magnetic field.
It utilizes null deflection -True
Null deflection is a crucial principle in paramagnetic analysers (reflected beam of light on two photocells) which gives very accurate results (typically 0.1%).It can be used to measure the concentration of diamagnetic gases – False
Since most other gases are weakly diamagnetic they are repelled by a magnetic field (nitric oxide is also paramagnetic).Can measure gases dissolved in the blood – False
For accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. Therefore, they are unsuitable to measure gases dissolved in blood.Does not require calibration – False
As with most measurement instruments paramagnetic analysers must be calibrated before use.E) The readings are unaffected by water vapour – False
Water vapour affects the readings hence for accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. That is why they are unsuitable to measure dissolved blood gases. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 19
Correct
-
A measuring system's response to change is complex, yet it can be mathematically modelled.
Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?
Your Answer: Dynamic second-order response
Explanation:The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:Zero-order:
Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.First-order:
Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.Second-order
Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 20
Correct
-
Regarding gas laws, which one best explains the ability of a Bourdon gauge to measure temperature?
Your Answer: Charles's law
Explanation:Charles’ Law states that there is a direct correlation between temperature and volume, where pressure and amount gas are constant. As temperature increases, volume also increases.
Boyle’s Law states that Pressure is inversely proportional to volume, assuming that temperature and amount of gas are constant. As volume increases, pressure decreases. In Dalton’s law of partial pressure, the total pressure exerted by a mixture of gases is equal to the sum of the partial pressure of the gases in mixture.
According to Henry’s Law for concentration of dissolved gases, at a constant temperature, the amount of a given gas that dissolves in a given type and volume of liquid is directly proportional to the partial pressure of that gas in equilibrium with that liquid. An equivalent way of stating the law is that the solubility of a gas in a liquid is directly proportional to the partial pressure of the gas above the liquid.
Gay-Lussac’s Law states that the pressure of a given mass of gas varies directly with the absolute temperature of the gas, when the volume is kept constant. This law is very similar to Charles’ Law, with the only difference being the type of container. Whereas the container in a Charles’ Law experiment is flexible, it is rigid in a Gay-Lussac’s Law experiment.
-
This question is part of the following fields:
- Basic Physics
-
00
Correct
00
Incorrect
00
:
00
:
00
Session Time
00
:
00
Average Question Time (
Mins)